Comparison of monitor unit calculations performed with a 3D computerized planning system and independent "hand" calculations: results of three years clinical experience.

التفاصيل البيبلوغرافية
العنوان: Comparison of monitor unit calculations performed with a 3D computerized planning system and independent "hand" calculations: results of three years clinical experience.
المؤلفون: Chan J; Department of Medical Physics, Hamilton Regional Cancer Centre, 699 Concession St., Hamilton, Ontario, Canada L8V 5C2. jackson.chan@hrcc.on.ca, Russell D, Peters VG, Farrell TJ
المصدر: Journal of applied clinical medical physics [J Appl Clin Med Phys] 2002 Autumn; Vol. 3 (4), pp. 293-301.
نوع المنشور: Comparative Study; Journal Article
اللغة: English
بيانات الدورية: Publisher: Wiley on behalf of American Association of Physicists in Medicine Country of Publication: United States NLM ID: 101089176 Publication Model: Print Cited Medium: Internet ISSN: 1526-9914 (Electronic) Linking ISSN: 15269914 NLM ISO Abbreviation: J Appl Clin Med Phys Subsets: MEDLINE
أسماء مطبوعة: Publication: 2017- : Malden, MA : Wiley on behalf of American Association of Physicists in Medicine
Original Publication: Reston, VA : American College of Medical Physics, c2000-
مواضيع طبية MeSH: Imaging, Three-Dimensional/*methods , Radiation Monitoring/*methods , Radiotherapy Planning, Computer-Assisted/*methods, Algorithms ; Brain/radiation effects ; Breast/radiation effects ; Cobalt Radioisotopes/therapeutic use ; Female ; Humans ; Male ; Prostate/radiation effects ; Radiation Monitoring/instrumentation ; Radiotherapy Dosage ; Radiotherapy Planning, Computer-Assisted/instrumentation ; Rectum/radiation effects
مستخلص: A comparison of the monitor unit calculations of a commercial 3D computerized treatment planning system (TPS) with "hand" calculations from lookup tables was made for a large number of clinical cases (greater than 13 500 treatment fields). Differences were analyzed by treatment site for prostate, rectum, cranium, and breast. The 3D TPS monitor unit calculation was systematically higher than the "hand" calculation by an amount that depended on the complexity of the treatment geometry. For simple geometries the mean difference was 1% and was as high as 3% for more complicated geometries. The higher value was attributed to an accumulation of differences introduced by multiple factors in the monitor unit calculation. Careful attention to factors such as patient contour could reduce the mean difference. "Hand" calculations were shown to be an accurate and useful tool for verification of TPS monitor unit calculations.
References: Phys Med Biol. 1998 Jun;43(6):1813-22. (PMID: 9651042)
J Appl Clin Med Phys. 2002 Spring;3(2):73-87. (PMID: 11958648)
Int J Radiat Oncol Biol Phys. 1993 May 20;26(2):261-73. (PMID: 8491684)
Med Phys. 1998 Oct;25(10):1773-829. (PMID: 9800687)
Phys Med Biol. 1972 Jan;17(1):42-51. (PMID: 5071500)
Med Phys. 1985 Mar-Apr;12(2):188-96. (PMID: 4000075)
J Appl Clin Med Phys. 2000 Summer;1(3):86-94. (PMID: 11674822)
Med Phys. 1993 Sep-Oct;20(5):1327-36. (PMID: 8289713)
J Appl Clin Med Phys. 2002 Winter;3(1):26-45. (PMID: 11818002)
Med Phys. 1999 Jul;26(7):1188-95. (PMID: 10435518)
J Appl Clin Med Phys. 2000 Autumn;1(4):120-5. (PMID: 11674827)
المشرفين على المادة: 0 (Cobalt Radioisotopes)
تواريخ الأحداث: Date Created: 20021018 Date Completed: 20021212 Latest Revision: 20200219
رمز التحديث: 20240829
مُعرف محوري في PubMed: PMC5724537
DOI: 10.1120/jacmp.v3i4.2553
PMID: 12383049
قاعدة البيانات: MEDLINE
الوصف
تدمد:1526-9914
DOI:10.1120/jacmp.v3i4.2553