دورية أكاديمية
Fracture of abutment screw supporting a cemented implant-retained prosthesis with external hexagon connection: a case report with sem evaluation.
العنوان: | Fracture of abutment screw supporting a cemented implant-retained prosthesis with external hexagon connection: a case report with sem evaluation. |
---|---|
المؤلفون: | Santos MD; Department of Implantology, University Santo Amaro, São Paulo, São Paulo, Brazil. marianabisi@hotmail.com, Pfeifer AB, Silva MR, Sendyk CL, Sendyk WR |
المصدر: | Journal of applied oral science : revista FOB [J Appl Oral Sci] 2007 Apr; Vol. 15 (2), pp. 148-51. |
نوع المنشور: | Journal Article |
اللغة: | English |
بيانات الدورية: | Publisher: Faculdade de Odontologia de Bauru, USP Country of Publication: Brazil NLM ID: 101189774 Publication Model: Print Cited Medium: Internet ISSN: 1678-7765 (Electronic) Linking ISSN: 16787757 NLM ISO Abbreviation: J Appl Oral Sci Subsets: PubMed not MEDLINE |
أسماء مطبوعة: | Original Publication: Bauru, SP : Faculdade de Odontologia de Bauru, USP, [2003]- |
مستخلص: | One of the causes of implant failures in cemented implant-retained prostheses is the fracture of abutment screw or UCLA abutment. This article reports a case of simultaneous fracture of two UCLA abutments screws occurring in an implant-supported prosthesis placed in the mandibular molar region. The fractured structures were examined under scanning electron microscopy to investigate the probable causes of the failure, which were not related to failures on materials or fabrication of the screws, but rather were due to shear forces. The misfit in cemented prostheses may be the most likely cause of shear force generation. |
References: | J Prosthet Dent. 1997 Jan;77(1):28-35. (PMID: 9029462) Int J Oral Maxillofac Implants. 2000 Sep-Oct;15(5):662-7. (PMID: 11055133) Int J Oral Maxillofac Implants. 1989 Fall;4(3):241-7. (PMID: 2700747) Int J Oral Maxillofac Implants. 2000 Jan-Feb;15(1):76-94. (PMID: 10697942) Int J Oral Surg. 1981 Dec;10(6):387-416. (PMID: 6809663) Int J Oral Maxillofac Implants. 1989 Summer;4(2):107-11. (PMID: 2599581) Int J Oral Maxillofac Implants. 2000 Jan-Feb;15(1):66-75. (PMID: 10697941) Int J Oral Maxillofac Implants. 1996 Mar-Apr;11(2):216-22. (PMID: 8666454) Int J Oral Maxillofac Implants. 2000 Mar-Apr;15(2):239-46. (PMID: 10795456) Int J Oral Maxillofac Implants. 1996 Sep-Oct;11(5):620-5. (PMID: 8908860) Int J Oral Maxillofac Implants. 2001 Mar-Apr;16(2):201-7. (PMID: 11324208) Int J Periodontics Restorative Dent. 1990;10(6):436-53. (PMID: 2098364) Int J Oral Maxillofac Implants. 2002 Jan-Feb;17(1):17-23. (PMID: 11858570) Pract Periodontics Aesthet Dent. 1995 Nov-Dec;7(9):15-8. (PMID: 9002890) |
تواريخ الأحداث: | Date Created: 20081218 Date Completed: 20100610 Latest Revision: 20211020 |
رمز التحديث: | 20240628 |
مُعرف محوري في PubMed: | PMC4327248 |
DOI: | 10.1590/s1678-77572007000200015 |
PMID: | 19089120 |
قاعدة البيانات: | MEDLINE |
كن أول من يترك تعليقا!