Efficacy of Orbscan II and Pentacam topographers by a repeatability analysis when assessing elevation maps in candidates to refractive surgery.

التفاصيل البيبلوغرافية
العنوان: Efficacy of Orbscan II and Pentacam topographers by a repeatability analysis when assessing elevation maps in candidates to refractive surgery.
المؤلفون: Núñez MX; Clínica de Oftalmología de Cali, Colombia. ximena@visionsana.com , Blanco C
المصدر: Biomedica : revista del Instituto Nacional de Salud [Biomedica] 2009 Sep; Vol. 29 (3), pp. 362-8.
نوع المنشور: Clinical Trial; Comparative Study; Journal Article; Research Support, Non-U.S. Gov't
اللغة: English
بيانات الدورية: Publisher: Instituto Nacional de Salud Country of Publication: Colombia NLM ID: 8205605 Publication Model: Print Cited Medium: Internet ISSN: 2590-7379 (Electronic) Linking ISSN: 01204157 NLM ISO Abbreviation: Biomedica Subsets: MEDLINE
أسماء مطبوعة: Publication: Bogota : Instituto Nacional de Salud
Original Publication: Bogotá, Colombia : El Instituto, [1981-
مواضيع طبية MeSH: Corneal Topography/*instrumentation , Keratoconus/*pathology, Adult ; Humans ; Middle Aged ; Refractive Surgical Procedures ; Reproducibility of Results ; Young Adult
مستخلص: Introduction: Anterior and posterior corneal elevations are measurements used to detect keratoconus suspects.
Purpose: To determine the efficacy of Orbscan II and Pentacam when assessing their elevation maps.
Materials and Methods: The efficacy of the Orbscan II and Pentacam measuring the anterior and posterior corneal elevations were evaluated in a sample of 68 eyes. The concordance between the two devices and the coefficient of repeatability were measured following the parameters of the British Standard Institution by the Bland-Altman concordance analysis and the Lin concordance correlation coefficient.
Results: The coefficient of repeatability at the point of maximum anterior elevation was 68.29% with the Orbscan and 24.20% with the Pentacam. The concordance correlation coefficient was 0.64 (CI 95%: 0.48-0.76) with the Orbscan and 0.94 with the Pentacam (CI 95%: 0.91-0.96). The coefficient of repeatability at the point of maximum posterior elevation was 38.7% with the Orbscan and 68.0% with the Pentacam. The concordance correlation coefficient was 0.69 with the Orbscan (CI 95%: 0.55-0.80) with a precision of 0.71 and an accuracy of 0.97, and 0.24 with the Pentacam (CI 95%: 0.00-0.45) with a precision of 0.24 and an accuracy of 0.99.
Conclusions: Measurement of the point of maximum posterior elevation is better with the Orbscan II and less precise with the Pentacam. The random error can be reduced by using the mean of three assessments and can serve as a guide in the search of diagnostic devices with minimum absolute relative error in all measurements.