دورية أكاديمية

A self-gating method for time-resolved imaging of nonuniform motion.

التفاصيل البيبلوغرافية
العنوان: A self-gating method for time-resolved imaging of nonuniform motion.
المؤلفون: Wundrak S; Department of Internal Medicine II, University Hospital of Ulm, Germany.; Sirona Dental Systems, Imaging Systems, Bensheim, Germany., Paul J; Department of Internal Medicine II, University Hospital of Ulm, Germany., Ulrici J; Sirona Dental Systems, Imaging Systems, Bensheim, Germany., Hell E; Sirona Dental Systems, Imaging Systems, Bensheim, Germany., Geibel MA; Department of Oral and Maxillofacial Surgery, University of Ulm, Germany., Bernhardt P; Department of Internal Medicine II, University Hospital of Ulm, Germany., Rottbauer W; Department of Internal Medicine II, University Hospital of Ulm, Germany., Rasche V; Department of Internal Medicine II, University Hospital of Ulm, Germany.
المصدر: Magnetic resonance in medicine [Magn Reson Med] 2016 Sep; Vol. 76 (3), pp. 919-25. Date of Electronic Publication: 2015 Oct 09.
نوع المنشور: Evaluation Study; Journal Article
اللغة: English
بيانات الدورية: Publisher: Wiley Country of Publication: United States NLM ID: 8505245 Publication Model: Print-Electronic Cited Medium: Internet ISSN: 1522-2594 (Electronic) Linking ISSN: 07403194 NLM ISO Abbreviation: Magn Reson Med Subsets: MEDLINE
أسماء مطبوعة: Publication: 1999- : New York, NY : Wiley
Original Publication: San Diego : Academic Press,
مواضيع طبية MeSH: Artifacts*, Arrhythmias, Cardiac/*diagnostic imaging , Cardiac-Gated Imaging Techniques/*methods , Image Enhancement/*methods , Image Interpretation, Computer-Assisted/*methods , Magnetic Resonance Imaging, Cine/*methods, Aged ; Aged, 80 and over ; Algorithms ; Female ; Humans ; Male ; Motion ; Reproducibility of Results ; Sensitivity and Specificity
مستخلص: Purpose: To develop a self-gating method capable of assessing nonuniform motion, e.g., in cardiovascular magnetic resonance imaging of patients with severe arrhythmia, or for imaging of the temporomandibular joint.
Methods: The proposed method allows cyclic motion trajectories with a nonuniform pace by replacing the one-dimensional gating signal of conventional image-based self-gating with a two-dimensional gating matrix. The resulting image quality is compared with conventional self-gating and real-time MRI.
Results: Nonuniform self-gating resulted in superior image quality compared with conventional self-gating and the feasibility study showed significantly improved image sharpness (P < 0.01). Further, improvements in image quality were shown compared with golden angle radial parallel sparse MRI.
Conclusion: A new self-gating method was proposed that allows cardiovascular magnetic resonance of arrhythmic patients, which is a common problem in clinical practice. Further, the proposed method enables self-gated imaging of the temporomandibular joint. Magn Reson Med 76:919-925, 2016. © 2015 Wiley Periodicals, Inc.
(© 2015 Wiley Periodicals, Inc.)
فهرسة مساهمة: Keywords: arrhythmia; cardiac; cardiovascular magnetic resonance; cine; self-gating; temporomandibular joint
تواريخ الأحداث: Date Created: 20151010 Date Completed: 20171218 Latest Revision: 20191210
رمز التحديث: 20221213
DOI: 10.1002/mrm.26000
PMID: 26451753