دورية أكاديمية

Quantifying osteosynthesis plate prominence - mathematical definitions and case study on a clavicle plate.

التفاصيل البيبلوغرافية
العنوان: Quantifying osteosynthesis plate prominence - mathematical definitions and case study on a clavicle plate.
المؤلفون: Zenker M; R&D Department, DePuy Synthes, Zuchwil, Switzerland., Theodorou E; R&D Department, DePuy Synthes, Zuchwil, Switzerland., Andermatt D; R&D Department, DePuy Synthes, Zuchwil, Switzerland., Winkler R; Innomedic GmbH, DePuy Synthes, Philippsburg-Rheinsheim, Germany., Hoyen HA; Department of Orthopaedic Surgery, MetroHealth Medical Center, Case Western Reserve University, Cleveland, OH, USA., Jaeger M; Department of Orthopedics and Trauma Surgery, Medical Center-Albert-Ludwigs-University of Freiburg, Freiburg, Germany., Jiang C; Shoulder Service, School of Medicine, Beijing Jishuitan Hospital, Peking University, Peking, China., Lambert S; Department of Trauma and Orthopedic Surgery, University College London Hospital NHS Foundation Trust, London, UK., Nijs S; Department of Trauma Surgery, University Hospitals Leuven, Leuven, Belgium., Fontana AD; R&D Department, DePuy Synthes, Zuchwil, Switzerland.
المصدر: Computer methods in biomechanics and biomedical engineering [Comput Methods Biomech Biomed Engin] 2023 Jun; Vol. 26 (8), pp. 917-926. Date of Electronic Publication: 2022 Jul 14.
نوع المنشور: Journal Article
اللغة: English
بيانات الدورية: Publisher: Informa Healthcare Country of Publication: England NLM ID: 9802899 Publication Model: Print-Electronic Cited Medium: Internet ISSN: 1476-8259 (Electronic) Linking ISSN: 10255842 NLM ISO Abbreviation: Comput Methods Biomech Biomed Engin Subsets: MEDLINE
أسماء مطبوعة: Publication: London : Informa Healthcare
Original Publication: [Amsterdam?] Netherlands : Gordon and Breach Science Publishers, c1997-
مواضيع طبية MeSH: Fractures, Bone*/surgery, Humans ; Clavicle/surgery ; Fracture Fixation, Internal/methods ; Bone Plates ; Retrospective Studies
مستخلص: Hardware prominence remains a clinical challenge in focus for implant design in subcutaneous plate applications. Existing evaluation of hardware prominence relies on plate-to-bone distance at a single point or on average. A reproducible measure for plate prominence remains undefined. This study mathematically defines the plate prominence linked to the cross-sectional area change due to the plate presence on the bone. Two anatomical plate designs were fitted to 100 clavicles, and afterwards plate prominence parameters were evaluated and compared. This methodology enables the quantification of hardware prominence for different plate designs to inform the development of implants targeting low prominence.
فهرسة مساهمة: Keywords: Plate prominence; clavicle fit; hardware prominence; maximum distance
تواريخ الأحداث: Date Created: 20220714 Date Completed: 20230505 Latest Revision: 20230505
رمز التحديث: 20240628
DOI: 10.1080/10255842.2022.2098676
PMID: 35833691
قاعدة البيانات: MEDLINE