Green, Validated HPLC Method Coupled With Fluorescence Detection for the Analysis of Tafluprost in its Pure Form and Ophthalmic Formulation: Application to Aqueous Humor and Content Uniformity Testing.

التفاصيل البيبلوغرافية
العنوان: Green, Validated HPLC Method Coupled With Fluorescence Detection for the Analysis of Tafluprost in its Pure Form and Ophthalmic Formulation: Application to Aqueous Humor and Content Uniformity Testing.
المؤلفون: Abd-AlGhafar WN; Mansoura University, Faculty of Pharmacy, Pharmaceutical Analytical Chemistry Department, Mansoura 35516, Egypt., Aly FA; Mansoura University, Faculty of Pharmacy, Pharmaceutical Analytical Chemistry Department, Mansoura 35516, Egypt., Sheribah ZA; Mansoura University, Faculty of Pharmacy, Pharmaceutical Analytical Chemistry Department, Mansoura 35516, Egypt., Saad S; Mansoura University, Faculty of Pharmacy, Pharmaceutical Analytical Chemistry Department, Mansoura 35516, Egypt.
المصدر: Journal of chromatographic science [J Chromatogr Sci] 2023 Feb 03; Vol. 61 (2), pp. 151-159.
نوع المنشور: Journal Article
اللغة: English
بيانات الدورية: Publisher: Oxford University Press Country of Publication: United States NLM ID: 0173225 Publication Model: Print Cited Medium: Internet ISSN: 1945-239X (Electronic) Linking ISSN: 00219665 NLM ISO Abbreviation: J Chromatogr Sci Subsets: MEDLINE
أسماء مطبوعة: Publication: <2012- > : Oxford : Oxford University Press
Original Publication: Niles Il : Preston Technical Abstracts
مواضيع طبية MeSH: Aqueous Humor* , Eye*, Chromatography, High Pressure Liquid/methods ; Ophthalmic Solutions
مستخلص: A green, simple and sensitive HPLC method coupled with fluorescence detection was implemented for the quantitative determination of the anti-glaucoma drug tafluprost (TFL). Liquid chromatography was performed on HyperClone™ ODS (C18) column of dimensions; 150 × 4.6 mm i.d. and 5 μm particle size using a green eluent; ethanol:0.01 M phosphate buffer (60:40 v/v, pH 4.5) delivered at 1 mL min-1. Fluorescence detection was accomplished at 220 nm (excitation) and 292 nm (emission). Bimatoprost (BIM) was used as an internal standard (I.S.). In this method, TFL was eluted after 6.70 minutes. The method satisfied International Council for Harmonization (ICH) validation guidelines, as proved by good linearity (r = 0.9999, over the range 0.05-2 μg mL-1), accuracy (recovery average 100.13 ± 1.27%), precision, robustness and specificity. The limit of detection and limit of quantification were found to be 0.016 and 0.048 μg mL-1, respectively. The proposed method has been successfully applied for the estimation of TFL in eye drops and aqueous humor. For the first time, the approach was applied with acceptable results for the evaluation of the uniformity of TFL eye drops content. Furthermore, Green Analytical Procedure Index (GAPI) and analytical Eco-scale were used to prove that the proposed HPLC method is environmentally friendly.
(© The Author(s) 2022. Published by Oxford University Press. All rights reserved. For permissions, please email: journals.permissions@oup.com.)
المشرفين على المادة: 1O6WQ6T7G3 (tafluprost)
0 (Ophthalmic Solutions)
تواريخ الأحداث: Date Created: 20220714 Date Completed: 20230208 Latest Revision: 20230208
رمز التحديث: 20230208
DOI: 10.1093/chromsci/bmac061
PMID: 35834929
قاعدة البيانات: MEDLINE