دورية أكاديمية

Use of transmission line analysis for multi-tuning of birdcage resonators.

التفاصيل البيبلوغرافية
العنوان: Use of transmission line analysis for multi-tuning of birdcage resonators.
المؤلفون: Pascone R; Department of Electrical Engineering, Manhattan College, Riverdale, NY 10471., Vullo T, Farrelly J, Mancuso R, Cahill PT
المصدر: Magnetic resonance imaging [Magn Reson Imaging] 1993; Vol. 11 (5), pp. 705-15.
نوع المنشور: Journal Article; Research Support, U.S. Gov't, P.H.S.
اللغة: English
بيانات الدورية: Publisher: Elsevier Country of Publication: Netherlands NLM ID: 8214883 Publication Model: Print Cited Medium: Print ISSN: 0730-725X (Print) Linking ISSN: 0730725X NLM ISO Abbreviation: Magn Reson Imaging Subsets: MEDLINE
أسماء مطبوعة: Publication: <2008->: Amsterdam : Elsevier
Original Publication: New York : Pergamon, c1982-
مواضيع طبية MeSH: Magnetic Resonance Imaging/*instrumentation, Technology, Radiologic
مستخلص: A general analysis for double- and higher order tuning of birdcage resonators is presented based on a lumped element transmission line model. Expressions were developed for the determination of the resonant frequencies of bandstop and bandpass birdcage coils and, with specific restrictions, for capacitor values required to obtain any two desired mode one resonant frequencies. Experimental measurements on three variants each of an eight-column bandstop and an eight-column bandpass were in excellent agreement with theory; the average absolute frequency difference and percent deviation were 1.51 +/- 1.57 MHz and 2.61 +/- 2.36%, respectively. In addition, 31P and 1H phantom images were obtained at 2.0 T using a representative bandstop coil.
معلومات مُعتمدة: 1 R15 CA49929-01A1 United States CA NCI NIH HHS
تواريخ الأحداث: Date Created: 19930101 Date Completed: 19930909 Latest Revision: 20190904
رمز التحديث: 20231215
DOI: 10.1016/0730-725x(93)90013-4
PMID: 8345785
قاعدة البيانات: MEDLINE
الوصف
تدمد:0730-725X
DOI:10.1016/0730-725x(93)90013-4