دورية أكاديمية

Gross Trunnion Failure in the Bipolar Hemiarthroplasty; Raising Concern about Short Trunnion: A Case Report.

التفاصيل البيبلوغرافية
العنوان: Gross Trunnion Failure in the Bipolar Hemiarthroplasty; Raising Concern about Short Trunnion: A Case Report.
المؤلفون: Kangbaek Kim, Junpyo Lee, Jeong Joon Yoo, Hee Joong Kim
المصدر: Hip & Pelvis; Mar2021, Vol. 33 Issue 1, p40-44, 5p
مستخلص: There have been some reports of gross trunnion failure (GTF) in total hip arthroplasty. Here, we report a case of GTF 19 years after bipolar hemiarthroplasty using a 28-mm head with a 14/16 taper bore. Compared to other GTF reports, the current case had some unusual aspects: bipolar hemiarthroplasty, 28-mm head, relatively lateonset, and no apparent findings of metallosis though a severe one was evident. A Computed tomography scout view provided valuable information in evaluating polyethylene, metal head, and neck inside the bipolar cup. The current report suggests a need for concerned regarding short trunnion length which may be associated with GTF. [ABSTRACT FROM AUTHOR]
Copyright of Hip & Pelvis is the property of Korean Hip Society and its content may not be copied or emailed to multiple sites or posted to a listserv without the copyright holder's express written permission. However, users may print, download, or email articles for individual use. This abstract may be abridged. No warranty is given about the accuracy of the copy. Users should refer to the original published version of the material for the full abstract. (Copyright applies to all Abstracts.)