A strategy to reduce the sensitivity of inhomogeneous magnetization transfer (ihMT) imaging to radiofrequency transmit field variations at 3 T.

التفاصيل البيبلوغرافية
العنوان: A strategy to reduce the sensitivity of inhomogeneous magnetization transfer (ihMT) imaging to radiofrequency transmit field variations at 3 T.
المؤلفون: Soustelle, Lucas, Troalen, Thomas, Hertanu, Andreea, Mchinda, Samira, Ranjeva, Jean‐Philippe, Guye, Maxime, Varma, Gopal, Alsop, David C., Duhamel, Guillaume, Girard, Olivier M.
المصدر: Magnetic Resonance in Medicine; Mar2022, Vol. 87 Issue 3, p1346-1359, 14p
مصطلحات موضوعية: MAGNETIZATION transfer, RADIO frequency, ROOT-mean-squares, WHITE matter (Nerve tissue)
مستخلص: Purpose: To minimize the sensitivity of inhomogeneous magnetization transfer gradient‐echo (ihMT‐GRE) imaging to radiofrequency (RF) transmit field (B1+) inhomogeneities at 3 T. Methods: The ihMT‐GRE sequence was optimized by varying the concentration of the RF saturation energy over time, obtained by increasing the saturation pulse power while extending the sequence repetition time (TR). Different protocols were tested using numerical simulations and human in vivo experiments in the brain white matter (WM) of healthy subjects at 3 T. The sensitivity of the ihMT ratio (ihMTR) to B1+ variations was investigated by comparing measurements obtained at nominal transmitter adjustments and following a 20% global B1+ drop. The resulting relative variations (δihMTR) were evaluated voxelwise as a function of the local B1+ distribution. The reproducibility of the protocol providing minimal B1+ bias was assessed in a test‐retest experiment. Results: In line with simulations, ihMT‐GRE experiments conducted at high concentration of the RF energy over time demonstrated strong reduction of the B1+ inhomogeneity effects in the human WM. Under the optimal conditions of 350‐ms TR and 3‐µT root mean square (RMS) saturation power, 73% of all WM voxels presented δihMTR below 10%. Reproducibility analysis yielded a close‐to‐zero systematic bias (ΔihMTR = −0.081%) and a high correlation (ρ² = 0.977) between test and retest experiments. Conclusion: Concentrating RF saturation energy in ihMT‐GRE sequences mitigates the sensitivity of the ihMTR to B1+ variations and allows for clinical‐ready ihMT imaging at 3 T. This feature is of particular interest for high and ultra‐high field applications. [ABSTRACT FROM AUTHOR]
Copyright of Magnetic Resonance in Medicine is the property of Wiley-Blackwell and its content may not be copied or emailed to multiple sites or posted to a listserv without the copyright holder's express written permission. However, users may print, download, or email articles for individual use. This abstract may be abridged. No warranty is given about the accuracy of the copy. Users should refer to the original published version of the material for the full abstract. (Copyright applies to all Abstracts.)
قاعدة البيانات: Complementary Index