A case of consistent discrepancies between urine and blood human chorionic gonadotropin measurements.

التفاصيل البيبلوغرافية
العنوان: A case of consistent discrepancies between urine and blood human chorionic gonadotropin measurements.
المؤلفون: Albersen, Arjan, Kemper-Proper, Evelien, Thelen, Marc H.M., Kianmanesh Rad, Noush A., Hoedemaeker, Robert F., Boesten, Lianne S.M.
المصدر: Clinical Chemistry & Laboratory Medicine; Jun2011, Vol. 49 Issue 6, p1029-1032, 4p, 2 Charts
مصطلحات موضوعية: CHORIONIC gonadotropins, URINALYSIS, BLOOD plasma, PITUITARY gland, POINT-of-care testing
مستخلص: Background: Our laboratory was confronted with two successive urine samples from a single patient which tested positive for human chorionic gonadotropin (hCG) when tested with both qualitative and quantitative assays, combined with no detectable hCG in corresponding plasma samples. Methods: Serial dilution and recovery experiments were performed in order to investigate the presence of interfering substances or a high-dose hook effect. The ovarian cysts that were removed from this patient were immunohistochemically stained using polyclonal anti-human hCG antibodies. Furthermore, a urine sample was sent to the USA hCG Reference Service for hCG variant analysis. Results: Dilution and recovery experiments in urine and plasma samples were unremarkable. The biopsy stained negative for human hCG and free β-subunit. hCG isoform analysis in the urine sample revealed that approximately 87.5% of the immunoreactive hCG lacked the β-subunit C-terminal peptide (CTP). Conclusions: We report a rare case in which two successive urine samples test positive for hCG whereas in corresponding plasma samples hCG is undetectable. The majority of the total hCG contained a degraded form of β-subunit that lacks the CTP. This hCG variant, possibly of pituitary origin, is thought to have an extreme fast clearance rate possibly explaining the discordance between the hCG results in urine and plasma samples. [ABSTRACT FROM AUTHOR]
Copyright of Clinical Chemistry & Laboratory Medicine is the property of De Gruyter and its content may not be copied or emailed to multiple sites or posted to a listserv without the copyright holder's express written permission. However, users may print, download, or email articles for individual use. This abstract may be abridged. No warranty is given about the accuracy of the copy. Users should refer to the original published version of the material for the full abstract. (Copyright applies to all Abstracts.)
قاعدة البيانات: Complementary Index
الوصف
تدمد:14346621
DOI:10.1515/CCLM.2011.158