A performance comparison of two kinds of liquid scintillation counters from PerkinElmer, Inc

التفاصيل البيبلوغرافية
العنوان: A performance comparison of two kinds of liquid scintillation counters from PerkinElmer, Inc
المؤلفون: Jian-hua Huang, Xiao-gui Feng, Guo-hua Jiang, Qian-ge He, Xuegang Liu, Jianchen Wang, Jian-yu Du, Jing Chen
المصدر: Journal of Radioanalytical and Nuclear Chemistry. 314:629-635
بيانات النشر: Springer Science and Business Media LLC, 2017.
سنة النشر: 2017
مصطلحات موضوعية: Light nucleus, Chemistry, Health, Toxicology and Mutagenesis, Counting efficiency, Liquid scintillation counting, Public Health, Environmental and Occupational Health, Analytical chemistry, 010403 inorganic & nuclear chemistry, 01 natural sciences, Pollution, Particle detector, 0104 chemical sciences, Analytical Chemistry, Nuclear Energy and Engineering, Performance comparison, Scintillation counter, Radiology, Nuclear Medicine and imaging, Spectroscopy
الوصف: The two kinds of liquid scintillation counters (LSCs) from PerkinElmer, Quantulus GCT 6220 and Quantulus 1220, have been compared on their basic performances, i.e., background count rate, counting efficiency, α/β discrimination, and instrument stability, by determining the radioactivity of the radionuclides of 3H, 14C, 241Am, 90Sr/90Y, 60Co, and 137Cs. Except for the highlight of Quantulus GCT 6220 in stability and that of Quantulus 1220 in background for α counting, the two kinds of LSCs are comparable in the basic performances to each other.
تدمد: 1588-2780
0236-5731
URL الوصول: https://explore.openaire.eu/search/publication?articleId=doi_________::959f9d535e85b54f7480e2209d1ba4f8
https://doi.org/10.1007/s10967-017-5428-2
حقوق: CLOSED
رقم الأكسشن: edsair.doi...........959f9d535e85b54f7480e2209d1ba4f8