Imaging performance of a large-field kinestatic charge detector for digital radiography |
المؤلفون: |
Senthilkumar Nagarajan, Carlos Rodriguez, Azad Mahmud, Herbert D. Zeman, Sreenivas Devidas, Shashidhar Kollipara, Joseph S. Laughter, George C. Giakos, Qian Peng, Frank A. DiBianca |
المصدر: |
SPIE Proceedings. |
بيانات النشر: |
SPIE, 1995. |
سنة النشر: |
1995 |
مصطلحات موضوعية: |
Detective quantum efficiency, Engineering, Data acquisition, Optics, business.industry, Optical engineering, Optical transfer function, Detector, X-ray detector, Electronic engineering, business, Image resolution, Digital radiography |
الوصف: |
The initial performance of a digital radiographic system incorporating a large-field (2016- channel) kinestatic charge detector and data acquisition electronics is discussed. The measured modulation transfer function of the system is 20% at 4 cy/mm. The measured detective quantum efficiency is 40 - 60%. These results are comparable with or better than those of current clinical (rare-earth film-screen and storage phosphor) systems. First images from the large-field system are shown and compared with those from commercial systems. Future system improvements in process or in planning are discussed.© (1995) COPYRIGHT SPIE--The International Society for Optical Engineering. Downloading of the abstract is permitted for personal use only. |