Slit camera focal spot measurement errors in mammography

التفاصيل البيبلوغرافية
العنوان: Slit camera focal spot measurement errors in mammography
المؤلفون: Gary T. Barnes, R. L. Tanner, S. Tang
المصدر: Medical Physics. 22:1803-1814
بيانات النشر: Wiley, 1995.
سنة النشر: 1995
مصطلحات موضوعية: Physics, Optics and Photonics, Microscope, Observational error, Observer (quantum physics), medicine.diagnostic_test, business.industry, X-Ray Film, Biophysics, Magnification, General Medicine, Biophysical Phenomena, law.invention, Optics, Tilt (optics), law, Calibration, medicine, Humans, Mammography, Female, Focal Spot Size, business, Technology, Radiologic
الوصف: Mammography x-ray tube focal spot sizes are routinely measured during acceptance testing and annual performance audits. The National Electrical Manufactures Association (NEMA) recommends the slit camera for this purpose. Investigated were the effect of slit rotational misalignment, tilt misalignment, image film density, film and screen-film image receptors, microscope magnification and reticule accuracy, and observer variation on slit camera focal spot measurements. Our results indicate that small rotational misalignment (< 5 degrees) and tilt misalignment (< 3 degrees) introduce insignificant error. Measured focal spot size increased slightly with image optical density, indicating that for consistent results the image optical density variations should be minimized. Also desirable for accurate field measurements is a high power microscope (25-50x) and a reticule with divisions of < or = 0.02 mm. Screen-film imaging consistently resulted in a slightly smaller measured focal spot size than direct film. The greatest source of error was due to observer variation. Of interest is that reader variability showed a consistent pattern and variation between two measurements by the same observer was much smaller than between observer variation, suggesting that standardized criteria should be established and a method of reader training developed. The length of the focal spot is defined at a reference axis angle specified by the mammography unit manufacturer. Presented is a tabulation of the focal spot geometry and reference axis angles for the majority of mammography units currently and recently marketed in North America.
تدمد: 0094-2405
URL الوصول: https://explore.openaire.eu/search/publication?articleId=doi_dedup___::496e6100068965802843d241f45442b1
https://doi.org/10.1118/1.597636
حقوق: CLOSED
رقم الأكسشن: edsair.doi.dedup.....496e6100068965802843d241f45442b1
قاعدة البيانات: OpenAIRE