A simple method for removing initial irregularity of an electrocardiogram during a transient state of a power supply in a defibrillator

التفاصيل البيبلوغرافية
العنوان: A simple method for removing initial irregularity of an electrocardiogram during a transient state of a power supply in a defibrillator
المؤلفون: Jeong-Han Yi, Jin-Su Ahn, Ki-Han Kim, Hyung Sik Kim
المصدر: Technology and Health Care
سنة النشر: 2020
مصطلحات موضوعية: Transient state, Computer science, 0206 medical engineering, Biomedical Engineering, Biophysics, Health Informatics, Bioengineering, 02 engineering and technology, electrocardiogram, Electric Capacitance, Electromagnetic interference, law.invention, galvanic isolation, Biomaterials, 03 medical and health sciences, Electrocardiography, Defibrillator, 0302 clinical medicine, Electric Power Supplies, law, Humans, Computer Simulation, Biosignal, irregularity, transient state, business.industry, Electrical engineering, Equipment Design, 020601 biomedical engineering, Power (physics), Capacitor, Transient (oscillation), Ground noise, business, Galvanic isolation, Electromagnetic Phenomena, 030217 neurology & neurosurgery, Information Systems, Defibrillators, Research Article
الوصف: BACKGROUND: The defibrillator is a device that instantaneously discharges the high energy stored in the capacitor to the human body to help revitalize the heart. The circuit for charging the capacitor uses the same power source as the biosignal measurement unit. Therefore, variation in main power supply voltage, ground noise, and electromagnetic interference from the charging circuit can induce distortion into the biosignal at the initial stage of charging. OBJECTIVE: In this study, a simple method is proposed for removing the initial irregularity of an electrocardiogram due to the transient state of a power supply. METHODS: To evaluate the method, a 1-channel electrocardiogram measurement unit and peripheral units were separated from the main control module using galvanic isolation. An isolated push-pull converter was designed to power the secondary side. The method was tested under steady-state and transient conditions. RESULTS: The obtained results proved that biosignal distortion can be significantly reduced. CONCLUSION: This method could be another simple implementation approach for solving signal distortions due to the transient status of power supplies used in medical devices.
تدمد: 1878-7401
URL الوصول: https://explore.openaire.eu/search/publication?articleId=doi_dedup___::52fbecd4c932512fc4049711a2f38e89
https://pubmed.ncbi.nlm.nih.gov/32364165
حقوق: OPEN
رقم الأكسشن: edsair.doi.dedup.....52fbecd4c932512fc4049711a2f38e89
قاعدة البيانات: OpenAIRE