العنوان: Positioning accuracy of a single-isocenter multiple targets SRS treatment: A comparison between Varian TrueBeam CBCT and Brainlab ExacTrac
المؤلفون: Samuel Kubler, Régis Ferrand, Elodie Martin, Eliane Graulieres
المصدر: Physica Medica. 80:267-273
بيانات النشر: Elsevier BV, 2020.
سنة النشر: 2020
مصطلحات موضوعية: Computer science, medicine.medical_treatment, Coordinate system, Biophysics, General Physics and Astronomy, Radiosurgery, Imaging phantom, 030218 nuclear medicine & medical imaging, 03 medical and health sciences, 0302 clinical medicine, Treatment plan, Calibration, medicine, Radiology, Nuclear Medicine and imaging, Computer vision, Image-guided radiation therapy, Phantoms, Imaging, business.industry, Truebeam, Isocenter, Spiral Cone-Beam Computed Tomography, General Medicine, Cone-Beam Computed Tomography, 030220 oncology & carcinogenesis, Artificial intelligence, Particle Accelerators, business
الوصف: Purpose This study compared the positioning accuracy between cone-beam CT (CBCT) and ExacTrac (ETX) for a single-isocenter multiple target stereotactic radiosurgery (SRS) on two TrueBeam STx systems. Methods A single-isocenter treatment plan was simulated on an anthropomorphic head phantom with six spherical steel ball bearings (BBs). One of the BBs was chosen to be the isocenter. The five off-isocenter targets were located at various distances from the isocenter. MV portal images were generated to evaluate the deviations between the expected and the real center of the targets after CBCT and ETX positioning, respectively. Results The evaluation of the positioning accuracy for the isocenter target showed that CBCT and ETX positioning provided comparable, sub-millimetric results. Deviations in positioning accuracy were also calculated for all other targets, also showing comparable results for CBCT and ETX. Moreover, our study showed that the deviation between CBCT and ETX positioning were in better agreement for TBSTx1 and deviated slightly higher on TBSTx2 (maximum: 1.23 mm at S/I direction), due to a less perfect alignment between the CBCT coordinate system and the ETX coordinate system on TBSTx2 compared to TBSTx1. This study also showed a correlation between the target positioning accuracy and the distance to the isocenter. Conclusion The positioning accuracy of ETX and CBCT for targets located at isocenter and off-isocenter locations was compared on two treatment machines and found comparable. Our study highlights the importance of a proper calibration procedure, to ensure correct alignment between the CBCT, ETX and machine coordinate systems.
تدمد: 1120-1797
URL الوصول: https://explore.openaire.eu/search/publication?articleId=doi_dedup___::7225b4d9c682a9b967df613d5cf08f99
https://doi.org/10.1016/j.ejmp.2020.10.022
حقوق: OPEN
رقم الأكسشن: edsair.doi.dedup.....7225b4d9c682a9b967df613d5cf08f99
قاعدة البيانات: OpenAIRE