العنوان: A New Proton CT Scanner
المؤلفون: Uzunyan, S. A., Blazey, G., Boi, S., Coutrakon, G., Dyshkant, A., Erdelyi, B., Hedin, D., Johnson, E., Krider, J., Rykalin, V., Zutshi, V., Ford, R., Sellberg, G., Rauch, J . E., Roman, M., Rubinov, P., Wilson, P., Naimuddin, M.
سنة النشر: 2014
المجموعة: Physics (Other)
مصطلحات موضوعية: Physics - Instrumentation and Detectors, Physics - Medical Physics
الوصف: The design, construction, and preliminary testing of a second generation proton CT scanner is presented. All current treatment planning systems at proton therapy centers use X-ray CT as the primary imaging modality for treatment planning to calculate doses to tumor and healthy tissues. One of the limitations of X-ray CT is in the conversion of X-ray attenuation coefficients to relative (proton) stopping powers, or RSP. This results in more proton range uncertainty, larger target volumes and therefore, more dose to healthy tissues. To help improve this, we present a novel scanner capable of high dose rates, up to 2~MHz, and large area coverage, 20~x~24~cm$^2$, for imaging an adult head phantom and reconstructing more accurate RSP values.
Comment: 6 pages,10 figures; Presented at 23rd Conference on Application of Accelerators in Research and Industry (CAARI 2014), San Antonio, TX, May 25-30, 2014
نوع الوثيقة: Working Paper
URL الوصول: http://arxiv.org/abs/1409.0049
رقم الأكسشن: edsarx.1409.0049
قاعدة البيانات: arXiv