دورية أكاديمية

Novel Dynamic Screw-Suture Stabilization System for Syndesmotic Repair Provides Better Anteroposterior Translation and Axial Tibiofibular Joint Stability: A Human Cadaveric Study

التفاصيل البيبلوغرافية
العنوان: Novel Dynamic Screw-Suture Stabilization System for Syndesmotic Repair Provides Better Anteroposterior Translation and Axial Tibiofibular Joint Stability: A Human Cadaveric Study
المؤلفون: Michael P. Swords DO, Firas Souleiman, Ivan Zderic, Torsten Pastor, Dominic Gehweiler, Jessica Galie, Todd J. Kent, Boyko Gueorguiev, Matthew P. Tomlinson FRCS(Orth), Tim Schepers MD
المصدر: Foot & Ankle Orthopaedics, Vol 7 (2022)
بيانات النشر: SAGE Publishing, 2022.
سنة النشر: 2022
المجموعة: LCC:Orthopedic surgery
مصطلحات موضوعية: Orthopedic surgery, RD701-811
الوصف: Category: Ankle; Basic Sciences/Biologics; Trauma Introduction/Purpose: The quest for optimal treatment of acute distal tibiofibular syndesmotic disruptions is still in full progress. Using suture-button repair devices is one of the dynamic stabilization options, however, they may not be always appropriate for stabilization of length-unstable syndesmotic injuries. Recently, a novel screw-suture repair system was developed to address such issues. The aim of this study was to investigate its performance in comparison to a suture-button stabilization of unstable syndesmotic injuries. Methods: Eight pairs of human cadaveric lower legs were CT scanned under 700 N single-leg axial loading in five foot positions - neutral, 15° external/internal rotation and 20° dorsi-/plantarflexion - in 3 different states: (1) pre-injured (intact); (2) injured, characterized by complete syndesmosis and deltoid ligaments cuts simulating pronation-eversion injury types III and IV as well as supination-eversion injury type IV according to Lauge-Hansen; (3) reconstructed, using a screw-suture (FIBULINK, Group 1) or a suture-button (TightRope, Group 2) implants for syndesmotic stabilization, placed 20 mm proximal to the tibia plafond/joint surface. Following, all specimens were: (1) biomechanically tested over 5000 cycles under combined 1400 N axial and +-15° torsional loading; (2) rescanned. Clear space (diastasis), anterior tibiofibular distance, talar dome angle and fibular shortening were measured radiologically from CT scans. Anteroposterior, axial, mediolateral and torsional movements at the distal tibiofibular joint level were evaluated biomechanically via motion tracking. Results: In each group clear space increased significantly after injury (p = 0.113). Conclusion: Although both implant systems demonstrate ability for stabilization of unstable syndesmotic injuries, the screw- suture reconstruction provides better anteroposterior translation and axial stability of the tibiofibular joint and maintains it over time under dynamic loading. Therefore, it could be considered as a valid option for treatment of syndesmotic disruptions.
نوع الوثيقة: article
وصف الملف: electronic resource
اللغة: English
تدمد: 2473-0114
24730114
Relation: https://doaj.org/toc/2473-0114
DOI: 10.1177/2473011421S00963
URL الوصول: https://doaj.org/article/29eec7adb4144440bd92136403dfe033
رقم الأكسشن: edsdoj.29eec7adb4144440bd92136403dfe033
قاعدة البيانات: Directory of Open Access Journals
الوصف
تدمد:24730114
DOI:10.1177/2473011421S00963