دورية أكاديمية

Comparison of Glidescope® Go™, King Vision™, Dahlhausen VL, I‑View™ and Macintosh laryngoscope use during difficult airway management simulation by experienced and inexperienced emergency medical staff: A randomized crossover manikin study.

التفاصيل البيبلوغرافية
العنوان: Comparison of Glidescope® Go™, King Vision™, Dahlhausen VL, I‑View™ and Macintosh laryngoscope use during difficult airway management simulation by experienced and inexperienced emergency medical staff: A randomized crossover manikin study.
المؤلفون: Andreas Moritz, Veronika Leonhardt, Johannes Prottengeier, Torsten Birkholz, Joachim Schmidt, Andrea Irouschek
المصدر: PLoS ONE, Vol 15, Iss 7, p e0236474 (2020)
بيانات النشر: Public Library of Science (PLoS), 2020.
سنة النشر: 2020
المجموعة: LCC:Medicine
LCC:Science
مصطلحات موضوعية: Medicine, Science
الوصف: BackgroundIn pre-hospital emergency care, video laryngoscopes (VLs) with disposable blades are preferably used due to hygienic reasons. However, there is limited existing data on the use of VLs with disposable blades by emergency medical staff. Therefore, the aim of this study was to compare the efficacy of four different VLs with disposable blades and the conventional standard Macintosh laryngoscope, when used by anesthetists with extensive previous experience and paramedics with little previous experience in endotracheal intubation (ETI) in a simulated difficult airway.MethodsFifty-eight anesthetists and fifty-four paramedics participated in our randomized crossover manikin trial. Each performed ETI with the new Glidescope® Go™, the Dahlhausen VL, the King Vision™, the I-View™ and the Macintosh laryngoscope. "Time to intubate" was the primary endpoint. Secondary endpoints were "time to vocal cords", "time to ventilate", overall success rate, number of intubation attempts and optimization maneuvers, Cormack-Lehane score, severity of dental compression and subjective impressions.ResultsThe Glidescope® Go™, the Dahlhausen VL and the King Vision™ provided superior intubation conditions in both groups without affecting the number of intubation attempts or the time required for successful intubation. When used by anesthetists with extensive experience in ETI, the use of VLs did not affect the overall success rate. In the hands of paramedics with little previous experience in ETI, the failure rate with the Macintosh laryngoscope (14.8%) decreased to 3.7% using the Glidescope® Go™ and the Dahlhausen VL. Despite the advantages of hyperangulated video laryngoscopes, the I-View™ performed worst.ConclusionsVLs with hyperangulated blades facilitated ETI in both groups and decreased the failure rate by an absolute 11.1% when used by paramedics with little previous experience in ETI. Our results therefore suggest that hyperangulated VLs could be beneficial and might be the method of choice in comparable settings, especially for emergency medical staff with less experience in ETI.
نوع الوثيقة: article
وصف الملف: electronic resource
اللغة: English
تدمد: 1932-6203
Relation: https://doaj.org/toc/1932-6203
DOI: 10.1371/journal.pone.0236474
URL الوصول: https://doaj.org/article/30192dbca5bb4e6fa7724c25435229cf
رقم الأكسشن: edsdoj.30192dbca5bb4e6fa7724c25435229cf
قاعدة البيانات: Directory of Open Access Journals
الوصف
تدمد:19326203
DOI:10.1371/journal.pone.0236474