Measurement of personal dose equivalent of X and gamma radiation by ring dosimeter: Results of intercomparison measurements

التفاصيل البيبلوغرافية
العنوان: Measurement of personal dose equivalent of X and gamma radiation by ring dosimeter: Results of intercomparison measurements
المؤلفون: Małgorzata Adamowicz, Sylwia Papierz, Janusz Kacprzyk, Zbigniew Kamiński, Marek Zmyślony
المصدر: Medycyna Pracy, Vol 64, Iss 5, Pp 631-637 (2013)
بيانات النشر: Nofer Institute of Occupational Medicine, 2013.
سنة النشر: 2013
المجموعة: LCC:Public aspects of medicine
مصطلحات موضوعية: interlaboratory comparisons, X-ray and gamma radiation, personal dose equivalent, ring dosimeter, thermoluminescence detectors, Public aspects of medicine, RA1-1270
الوصف: Background: The aim of the study was to present the results of the interlaboratory comparisons for ring dosimeters and to confirm that the applied method is suitable for measuring the personal dose equivalent HP(0.07). In addition, calibration procedures used in dosimetric measurements in persons occupationally exposed to ionizing photon radiation X and γ were presented. Materials and Methods: Ring dosimeters made of flexible plastics with the diameter of approximately 20 mm, equipped with two thermoluminescence (TL) detectors type MTS-N, were the subject of interlaboratory comparisons. Irradiated detectors were red out using a new manual TLD's reader (FIMEL, France). All TLD exposures were done for validation of TLD readers and were performed using the reference X-ray and γ beams with the ISO rod phantom. The methodology of performed exposures corresponded with the methods applied by the European Radiation Dosimetry Group (EURADOS), the European organizer of interlaboratory comparisons. Results: The energy, dose and angular characteristics of the ring dosimeter allowed to elaborate the formula for estimating and verifying the personal dose equivalent HP(0.07). The test was performed to check and confirm the correctness of the estimated characteristics. The test results were satisfactory, and thus the readiness to implement TLD reader in the used method and to participate in interlaboratory comparisons was confirmed. Conclusions: According to the requirements of the Polish Centre for Accreditation, the laboratory was participating in the interlaboratory comparison organized by EURADOS in terms of the personal dose equivalent HP(0.07). The result of the comparison was satisfactory, therefore the correctness of the testing procedure was confirmed. Med Pr 2013;64(5):631–637
نوع الوثيقة: article
وصف الملف: electronic resource
اللغة: English
Polish
تدمد: 0465-5893
2353-1339
Relation: http://medpr.imp.lodz.pl/Pomiar-indywidualnego-rownowaznika-dawki-promieniowania-X-i-gamma-za-pomoca-dozymetru-pierscionkowego-wyniki-porownan-miedzylaboratoryjnych,498,0,2.html; https://doaj.org/toc/0465-5893; https://doaj.org/toc/2353-1339
DOI: 10.13075/mp.5893.2013.0055
URL الوصول: https://doaj.org/article/c7db6a20983447f4ac9e98ba850cc91f
رقم الأكسشن: edsdoj.7db6a20983447f4ac9e98ba850cc91f
قاعدة البيانات: Directory of Open Access Journals