دورية أكاديمية

[Catheter fracture and embolization: a rare complication of a permanent implanted intravenous catheter system].

التفاصيل البيبلوغرافية
العنوان: [Catheter fracture and embolization: a rare complication of a permanent implanted intravenous catheter system].
عنوان ترانسليتريتد: Katheterbreuk en -embolisatie: een zeldzame complicatie van een langdurig geïmplanteerd intraveneus kathetersysteem.
المؤلفون: Mazel JW; Medisch Centrum Haaglanden, afd. Heelkunde, Leidschendam., Idenburg FJ, van Delden OM
المصدر: Nederlands tijdschrift voor geneeskunde [Ned Tijdschr Geneeskd] 2000 Jul 08; Vol. 144 (28), pp. 1360-3.
نوع المنشور: Case Reports; Journal Article
اللغة: Dutch; Flemish
بيانات الدورية: Publisher: Vereniging NTvG Country of Publication: Netherlands NLM ID: 0400770 Publication Model: Print Cited Medium: Print ISSN: 0028-2162 (Print) Linking ISSN: 00282162 NLM ISO Abbreviation: Ned Tijdschr Geneeskd Subsets: MEDLINE
أسماء مطبوعة: Publication: 2017?-: Amsterdam : Vereniging NTvG
Original Publication: Houten : Bohn Stafleu van Loghum
مواضيع طبية MeSH: Antineoplastic Agents/*administration & dosage , Breast Neoplasms/*drug therapy , Catheterization, Central Venous/*adverse effects , Catheters, Indwelling/*adverse effects , Embolism/*etiology, Breast Neoplasms/pathology ; Catheterization, Central Venous/methods ; Embolism/diagnostic imaging ; Equipment Failure ; Female ; Humans ; Middle Aged ; Radiography, Thoracic ; Reoperation ; Subclavian Vein/diagnostic imaging
مستخلص: An implantable venous access system was used in a 55-year-old woman with metastatic breast cancer for the delivery of chemotherapy. Four months after implantation the catheter was resistant to the injection of fluids. A chest X-ray showed fracture of the catheter with embolisation to the right pulmonary artery. Analysis of the fractured catheter after removal showed that the fracture was caused by catheter pinch-off. Catheter pinch-off is caused by friction of the catheter between the clavicle and the first rib. The incidence of this rare complication is estimated at 0.1-1%. The incidence of catheter pinch-off can be reduced by a lateral insertion technique and by radiographic monitoring after implantation.
المشرفين على المادة: 0 (Antineoplastic Agents)
تواريخ الأحداث: Date Created: 20000803 Date Completed: 20000825 Latest Revision: 20161124
رمز التحديث: 20240627
PMID: 10923159
قاعدة البيانات: MEDLINE