دورية أكاديمية

[Measurement artifacts in Gaeltec intracranial pressure monitors due to radio waves from personal beeper systems].

التفاصيل البيبلوغرافية
العنوان: [Measurement artifacts in Gaeltec intracranial pressure monitors due to radio waves from personal beeper systems].
عنوان ترانسليتريتد: Messartefakte bei Gaeltec-Hirndrucksonden durch Radiowellen von Personensuchanlagen.
المؤلفون: Betsch HM; Neurochirurgische Universitätsklinik Heidelberg., Aschoff A
المصدر: Anasthesiologie, Intensivmedizin, Notfallmedizin, Schmerztherapie : AINS [Anasthesiol Intensivmed Notfallmed Schmerzther] 1992 Feb; Vol. 27 (1), pp. 51-2.
نوع المنشور: English Abstract; Journal Article
اللغة: German
بيانات الدورية: Publisher: Thieme Country of Publication: Germany NLM ID: 9109478 Publication Model: Print Cited Medium: Print ISSN: 0939-2661 (Print) Linking ISSN: 09392661 NLM ISO Abbreviation: Anasthesiol Intensivmed Notfallmed Schmerzther Subsets: MEDLINE
أسماء مطبوعة: Original Publication: Stuttgart ; New York : Thieme, c1991-
مواضيع طبية MeSH: Artifacts* , Hospital Communication Systems* , Intracranial Pressure* , Radio Waves*, Monitoring, Physiologic/*instrumentation, Humans
مستخلص: Artifacts were observed when using epidural pressure monitoring systems (Gaeltec ICT/B Titan). These were produced by radio waves from our beeper system. A non-existing electromagnetic shield of the Gaeltec-measuring amplifier SN 10/11 could be found as the cause. The artifacts can lead to therapeutical errors.
تواريخ الأحداث: Date Created: 19920201 Date Completed: 19920924 Latest Revision: 20061115
رمز التحديث: 20221213
DOI: 10.1055/s-2007-1000251
PMID: 1504191
قاعدة البيانات: MEDLINE
الوصف
تدمد:0939-2661
DOI:10.1055/s-2007-1000251