A comparison of applanation tonometry using conventional reusable goldmann prisms and disposable prisms.

التفاصيل البيبلوغرافية
العنوان: A comparison of applanation tonometry using conventional reusable goldmann prisms and disposable prisms.
المؤلفون: Tsai AS; Singapore National Eye Centre †Department of Ophthalmology, National University of Singapore, Singapore., How AC, Su DH, Lee KY, Wong TT, Perera SA
المصدر: Journal of glaucoma [J Glaucoma] 2014 Oct-Nov; Vol. 23 (8), pp. 521-5.
نوع المنشور: Journal Article; Observational Study
اللغة: English
بيانات الدورية: Publisher: Wolters Kluwer Health, Inc Country of Publication: United States NLM ID: 9300903 Publication Model: Print Cited Medium: Internet ISSN: 1536-481X (Electronic) Linking ISSN: 10570829 NLM ISO Abbreviation: J Glaucoma Subsets: MEDLINE
أسماء مطبوعة: Publication: <2015- > : Philadelphia, PA : Wolters Kluwer Health, Inc.
Original Publication: New York, N.Y. : Raven Press, c1992-
مواضيع طبية MeSH: Glaucoma/*diagnosis , Intraocular Pressure/*physiology , Tonometry, Ocular/*instrumentation, Adult ; Aged ; Aged, 80 and over ; Disposable Equipment ; Female ; Glaucoma/physiopathology ; Humans ; Male ; Middle Aged ; Prospective Studies ; Retrospective Studies ; Tonometry, Ocular/methods ; Young Adult
مستخلص: Purpose: To determine the agreement between intraocular pressure (IOP) measurements using conventional Goldmann applanation tonometry (GAT) and Tonosafe disposable prisms, and also to provide a comprehensive cost analysis of the use of both types of prisms
Methods: : In this prospective observational study, 198 eyes of 100 glaucoma patients had their IOPs measured by 5 consultant ophthalmologists. Data were analyzed using the Bland-Altman method of differences, and correlation was measured using the Pearson coefficient. An analysis of the cost incurred using the 2 methods over a 6-month period was performed.
Results: The majority were Chinese (82%), with a male preponderance (57%). The range of IOPs as measured by GAT was 4 to 34 mm Hg. Using the Bland-Altman method to compare GAT and disposable prisms, the bias was 0.2 mm Hg. Tonosafe overestimated the IOP by 0.2 mm Hg in the right eye and underestimated it by 0.2 mm Hg in the left eye. The Tonosafe IOP correlated well with GAT, with a Pearson coefficient of correlation(r) of 0.91 (P<0.0005) for the right eye and 0.92 (P<0.0005) for the left eye, respectively. For those with GAT IOP≥21 mm Hg (n=26), Tonosafe underestimated the IOP by 0.35 mm Hg. The cost incurred by Tonosafe prisms was approximately 8 times that of GAT, but the cost differential reverses when GAT had to be replaced after every 100 cycles of disinfection.
Conclusions: We found a good correlation between Tonosafe prisms and conventional GAT in measuring the IOP. Tonosafe prisms may be of use, especially if the risk of transmission of infection is high. However, cost may limit its more widespread use.
تواريخ الأحداث: Date Created: 20130502 Date Completed: 20150207 Latest Revision: 20141002
رمز التحديث: 20240628
DOI: 10.1097/IJG.0b013e318294863d
PMID: 23632407