دورية أكاديمية

[CONTINUOUS HEMOGLOBIN MONITORING USING PULSE CO-OXIMETRY IN CARDIAC SURGERY.]

التفاصيل البيبلوغرافية
العنوان: [CONTINUOUS HEMOGLOBIN MONITORING USING PULSE CO-OXIMETRY IN CARDIAC SURGERY.]
المؤلفون: Len'kin PI, Smetkin AA, Husseyn A, Fot EV, Len'kin AI, Paromov KV, Ushakov AA, Krygina MA, Kirov MY
المصدر: Anesteziologiia i reanimatologiia [Anesteziol Reanimatol] 2016 Sep; Vol. 61 (5), pp. 329-334.
نوع المنشور: Journal Article
اللغة: Russian
بيانات الدورية: Publisher: Meditsina Country of Publication: Russia (Federation) NLM ID: 7705399 Publication Model: Print Cited Medium: Print ISSN: 0201-7563 (Print) Linking ISSN: 02017563 NLM ISO Abbreviation: Anesteziol Reanimatol Subsets: MEDLINE
أسماء مطبوعة: Publication: Moskva : Meditsina
Original Publication: Moskva, Meditsina.
مواضيع طبية MeSH: Cardiac Surgical Procedures* , Oximetry*, Hemoglobins/*analysis , Monitoring, Physiologic/*methods, Aged ; Female ; Humans ; Male ; Middle Aged ; Prospective Studies ; Sensitivity and Specificity
مستخلص: The Aim: To assess the accuracy ofcontinuous hemoglobin monitoring using pulse co-oximetry and revealfactors affecting the results of the measurements during early postoperative period in cardiac surgery.
Materials and Methods: 27 patients undergoing off-pump CABG and 16patients after elective complex (repair or replacement of two or more valves) or combined (valve and coronary artery) cardiac surgery requiring CPB were enrolled into a prospective observational study. Both groups received continuous hemoglobin monitoring using pulse co-oximetry (SpHb). During early postoperative period SpHb was compared with hemoglobin concentration in the arterial blood (Hbart).
Results: Wefoundpositive correlation between SpHb and Hb in both groups (rho =0,29, p < 0,05 u rho=0,34; p<0.005 respectively). The Bland-Altman analysis showed a bias ? limits of agreement (?I.96 SD) between the continuously measured hemoglobin and reference arterial blood hemoglobin concentration of -6,0 ? 41,0 g/l in the off-pump group and 7,7 k 31,0 g/l in the CPB group.
Conclusions: Hemoglobin measurement using pulse co-oximetry does not provide acceptable accuracy during early postoperative period in cardiac surgery. Applicability of this technology is influenced by vascular tone, systemic and regional tissue hypoperfusion.
المشرفين على المادة: 0 (Hemoglobins)
تواريخ الأحداث: Date Created: 20180301 Date Completed: 20180327 Latest Revision: 20180327
رمز التحديث: 20231215
PMID: 29489097
قاعدة البيانات: MEDLINE