دورية أكاديمية

Comparison between pulse wave velocities measured using Complior and measured using Biopac.

التفاصيل البيبلوغرافية
العنوان: Comparison between pulse wave velocities measured using Complior and measured using Biopac.
المؤلفون: van Velzen MHN; Laboratory of Experimental Anesthesiology, Department of Anesthesiology, Erasmus University Medical Center, Rotterdam, The Netherlands. m.v.velzen@jbz.nl.; Department of Medical Information Communication Technology MICT, Jeroen Bosch Ziekenhuis, PO Box 90153, 5200 ME, 's Hertogenbosch, The Netherlands. m.v.velzen@jbz.nl., Stolker RJ; Laboratory of Experimental Anesthesiology, Department of Anesthesiology, Erasmus University Medical Center, Rotterdam, The Netherlands., Loeve AJ; Department of BioMechanical Engineering, Faculty 3mE, Delft University of Technology, Delft, The Netherlands., Niehof SP; Laboratory of Experimental Anesthesiology, Department of Anesthesiology, Erasmus University Medical Center, Rotterdam, The Netherlands., Mik EG; Laboratory of Experimental Anesthesiology, Department of Anesthesiology, Erasmus University Medical Center, Rotterdam, The Netherlands.
المصدر: Journal of clinical monitoring and computing [J Clin Monit Comput] 2019 Apr; Vol. 33 (2), pp. 241-247. Date of Electronic Publication: 2018 Jun 06.
نوع المنشور: Comparative Study; Journal Article
اللغة: English
بيانات الدورية: Publisher: Springer Country of Publication: Netherlands NLM ID: 9806357 Publication Model: Print-Electronic Cited Medium: Internet ISSN: 1573-2614 (Electronic) Linking ISSN: 13871307 NLM ISO Abbreviation: J Clin Monit Comput
أسماء مطبوعة: Publication: Amsterdam : Springer
Original Publication: Dordrecht, The Netherlands ; Boston : Kluwer Academic Publishers, c1998-
مواضيع طبية MeSH: Blood Flow Velocity* , Pulse Wave Analysis* , Vascular Stiffness*, Monitoring, Physiologic/*instrumentation, Adult ; Cardiovascular Diseases ; Female ; Femoral Artery ; Humans ; Male ; Monitoring, Physiologic/methods ; Radial Artery ; Risk ; Software ; Young Adult
مستخلص: Arterial stiffness is a reliable prognostic parameter for cardiovascular diseases. The effect of change in arterial stiffness can be measured by the change of the pulse wave velocity (PWV). The Complior system is widely used to measure PWV between the carotid and radial arteries by means of piezoelectric clips placed around the neck and the wrist. The Biopac system is an easier to use alternative that uses ECG and simple optical sensors to measure the PWV between the heart and the fingertips, and thus extends a bit more to the peripheral vasculature compared to the Complior system. The goal of this study was to test under various conditions to what extent these systems provide comparable and correlating values. 25 Healthy volunteers, 20-30 years old, were measured in four sequential position: sitting, lying, standing and sitting. The results showed that the Biopac system measured consistently and significantly lower PWV values than the Complior system, for all positions. Correlation values and Bland-Altman plots showed that despite the difference in PWV magnitudes obtained by the two systems the measurements did agree well. Which implies that as long as the differences in PWV magnitudes are taken into account, either system could be used to measure PWV changes over time. However, when basing diagnosis on absolute PWV values, one should be very much aware of how the PWV was measured and with what system.
References: Circulation. 1999 May 11;99(18):2434-9. (PMID: 10318666)
Am J Physiol. 1999 Dec;277(6 Pt 2):S234-43. (PMID: 10644250)
Hypertension. 2001 May;37(5):1236-41. (PMID: 11358934)
Circulation. 2002 Oct 15;106(16):2085-90. (PMID: 12379578)
Arch Mal Coeur Vaiss. 2002 Dec;95(12):1215-8. (PMID: 12611043)
Diabet Med. 2004 Mar;21(3):262-6. (PMID: 15008837)
J Hypertens. 2008 Oct;26(10):2001-7. (PMID: 18806624)
J Hypertens. 2009 Nov;27(11):2186-91. (PMID: 19834344)
Psychophysiology. 1979 Nov;16(6):546-53. (PMID: 229507)
Physiol Meas. 2016 Jan;37(1):52-66. (PMID: 26641656)
Med Biol Eng Comput. 2017 Nov;55(11):1989-2000. (PMID: 28361357)
PLoS One. 2018 Oct 10;13(10):e0204105. (PMID: 30304059)
Circulation. 1968 Feb;37(2):149-59. (PMID: 5640345)
Circulation. 1983 Jul;68(1):50-8. (PMID: 6851054)
Hypertension. 1995 Sep;26(3):485-90. (PMID: 7649586)
J Hypertens. 1998 Dec;16(12 Pt 2):2079-84. (PMID: 9886900)
فهرسة مساهمة: Keywords: Arterial stiffness; Cardiovascular disease; Non-invasive; Pulse wave velocity
تواريخ الأحداث: Date Created: 20180608 Date Completed: 20190819 Latest Revision: 20200225
رمز التحديث: 20240628
مُعرف محوري في PubMed: PMC6420418
DOI: 10.1007/s10877-018-0165-9
PMID: 29876710
قاعدة البيانات: MEDLINE
الوصف
تدمد:1573-2614
DOI:10.1007/s10877-018-0165-9