دورية أكاديمية

Calculating equivalent dose received from a patient undergoing nuclear medicine procedure by merge phantoms tool and GAMOS/Geant4 6.0.0 software.

التفاصيل البيبلوغرافية
العنوان: Calculating equivalent dose received from a patient undergoing nuclear medicine procedure by merge phantoms tool and GAMOS/Geant4 6.0.0 software.
المؤلفون: Thao NTP; Vietnam Atomic Energy Institute, 59 Ly Thuong Kiet, Tran Hung Dao, Hoan Kiem, Ha Noi, Viet Nam; Nguyen Huu Huan High School, 1 Doan Ket, Binh Tho, Thu Duc, Ho Chi Minh City, Viet Nam. Electronic address: thi.phuong.thao.8488@gmail.com., Trung NT; TechBase, 67 Le Loi, Ben Nghe, Distric 1, Ho Chi Minh City, Viet Nam., Son ND; VNUHCM-University of Science, 227 Nguyen Van Cu, District 5, Ho Chi Minh City, Viet Nam.
المصدر: Physica medica : PM : an international journal devoted to the applications of physics to medicine and biology : official journal of the Italian Association of Biomedical Physics (AIFB) [Phys Med] 2019 Dec; Vol. 68, pp. 41-46. Date of Electronic Publication: 2019 Nov 14.
نوع المنشور: Journal Article
اللغة: English
بيانات الدورية: Publisher: Istituti Editoriali e Poligrafici Internazionali Country of Publication: Italy NLM ID: 9302888 Publication Model: Print-Electronic Cited Medium: Internet ISSN: 1724-191X (Electronic) Linking ISSN: 11201797 NLM ISO Abbreviation: Phys Med Subsets: MEDLINE
أسماء مطبوعة: Publication: Pisa : Istituti Editoriali e Poligrafici Internazionali
Original Publication: Lugano, Switzerland : Giardini editori S.A.,
مواضيع طبية MeSH: Monte Carlo Method* , Nuclear Medicine* , Phantoms, Imaging* , Radiation Dosage* , Software*, Positron Emission Tomography Computed Tomography/*instrumentation, Humans
مستخلص: Purpose: This report introduces a tool for merging two voxel phantoms to calculate the deposited dose that a person receives from a patient undergoing Nuclear medicine procedures.
Materials and Methods: The phantoms must be converted to the text format used by GEANT4 to treat DICOM images via the GAMOS utilities. The Merge Phantoms Tool can merge two phantoms in two different cases: standing either side by side or opposite. The merged phantom is also in text format and is subsequently input back into GAMOS to calculate the equivalent dose that a person receives from a patient. The equivalent doses to the eyes of people in contact are calculated in a case where a patient was administered 185 MBq of 18F-FDG during a PET examination.
Results: The corresponding doses when the two phantoms are standing opposite are greater than those when they are standing side by side and smaller than those from point and tube source calculated by Sumi Yokoyama at any distance.
Conclusions: The Merge phantoms tool and GAMOS software can be used to calculate the deposited dose that a person receives from a patient. An accurate dose calculation can be used for radiation protection, or deciding whether a patient can be released from isolation if the dose is small even in a close contact.
(Copyright © 2019 Associazione Italiana di Fisica Medica. Published by Elsevier Ltd. All rights reserved.)
فهرسة مساهمة: Keywords: DICOM; Equivalent dose; GAMOS; GEANT4; Nuclear medicine; Phantom merging
تواريخ الأحداث: Date Created: 20191117 Date Completed: 20200428 Latest Revision: 20200428
رمز التحديث: 20221213
DOI: 10.1016/j.ejmp.2019.11.004
PMID: 31733405
قاعدة البيانات: MEDLINE