دورية أكاديمية

Steam Sterilization Chemical Indicators Are Not Adequate for Monitoring Real Steam Sterilization Cycles.

التفاصيل البيبلوغرافية
العنوان: Steam Sterilization Chemical Indicators Are Not Adequate for Monitoring Real Steam Sterilization Cycles.
المؤلفون: Laranjeira PR; Cardinal Health, 7000 Cardinal Place, Dublin, OH 43017; and prlaranjeira@gmail.com., de Souza RQ; Nursing School (EEUSP), University of São Paulo, Brazil., Bronzatti JAG; Nursing School (EEUSP), University of São Paulo, Brazil., Graziano KU; Nursing School (EEUSP), University of São Paulo, Brazil.
المصدر: PDA journal of pharmaceutical science and technology [PDA J Pharm Sci Technol] 2020 Jul-Aug; Vol. 74 (4), pp. 435-438.
نوع المنشور: Comparative Study; Journal Article; Research Support, Non-U.S. Gov't
اللغة: English
بيانات الدورية: Publisher: Pda (Parenteral Drug Association) Country of Publication: United States NLM ID: 9439538 Publication Model: Print Cited Medium: Internet ISSN: 1948-2124 (Electronic) Linking ISSN: 10797440 NLM ISO Abbreviation: PDA J Pharm Sci Technol Subsets: MEDLINE
أسماء مطبوعة: Original Publication: Bethesda Md : Pda (Parenteral Drug Association)
مواضيع طبية MeSH: Decontamination* , Steam* , Sterilization*, Equipment Contamination/*prevention & control , Indicators and Reagents/*chemistry, Hospitals ; Temperature ; Time Factors
مستخلص: Chemical indicators are commonly used in hospitals to monitor steam sterilization conditions, indicating that medical devices are safe to be used. The results are stored for future evidence in the event of an infection incident root cause analysis. This type of indicator is also becoming an option for cycle monitoring in pharmaceutical steam sterilizers, improving cycle control. They are constructed and tested according to published standards, but contradictory results between chemical indicators and cycle printouts have a critical impact on process control. We found that Type 6 chemical indicators used in steam sterilizer cycles did not perform according to their intended use, showing an "approved" result in a "failed" cycle (a false positive). This study demonstrated that Type 6 chemical indicator specifications are not adequate for monitoring steam sterilizers. A change in standards is therefore needed.
(© PDA, Inc. 2020.)
فهرسة مساهمة: Keywords: Biological indicators; Chemical indicators; Cycle monitoring; Process indicators; Steam sterilization, Medical device
المشرفين على المادة: 0 (Indicators and Reagents)
0 (Steam)
تواريخ الأحداث: Date Created: 20200802 Date Completed: 20210602 Latest Revision: 20210602
رمز التحديث: 20231215
DOI: 10.5731/pdajpst.2019.09886
PMID: 32737242
قاعدة البيانات: MEDLINE
الوصف
تدمد:1948-2124
DOI:10.5731/pdajpst.2019.09886