دورية أكاديمية
Influence of Flat Occlusal Splint on Stresses Induced on Implants for Different Fixed Prosthetic Systems.
العنوان: | Influence of Flat Occlusal Splint on Stresses Induced on Implants for Different Fixed Prosthetic Systems. |
---|---|
المؤلفون: | Henrique MN; Postgraduate Program, School of Dentistry, University of Taubaté., Caldas RA; Postgraduate Program, School of Dentistry, State University of Campinas., Baroudi K; Postgraduate Program, School of Dentistry, University of Taubaté., Amaral M; Postgraduate Program, School of Dentistry, University of Taubaté., Vitti RP; Postgraduate Program, School of Dentistry, University of Taubaté., Silva-Concílio LR; Postgraduate Program, School of Dentistry, University of Taubaté. |
المصدر: | The European journal of prosthodontics and restorative dentistry [Eur J Prosthodont Restor Dent] 2021 May 28; Vol. 29 (2), pp. 84-92. Date of Electronic Publication: 2021 May 28. |
نوع المنشور: | Journal Article |
اللغة: | English |
بيانات الدورية: | Publisher: Stephen Hancocks Ltd. in association with Dennis Barber Ltd Country of Publication: England NLM ID: 9314899 Publication Model: Electronic Cited Medium: Print ISSN: 0965-7452 (Print) Linking ISSN: 09657452 NLM ISO Abbreviation: Eur J Prosthodont Restor Dent |
أسماء مطبوعة: | Publication: 2012- : Lowestoft : Stephen Hancocks Ltd. in association with Dennis Barber Ltd. Original Publication: Larkfield, Aylesford, Kent, UK : Mosby-Year Book Europe Ltd., c1992- |
مواضيع طبية MeSH: | Dental Implants* , Dental Prosthesis, Implant-Supported*, Computer Simulation ; Dental Prosthesis Design ; Dental Stress Analysis ; Finite Element Analysis ; Occlusal Splints ; Stress, Mechanical |
مستخلص: | Objective: The flat occlusal plate has been recommended to reduce stress concentration in implant prosthesis treatments. The purpose was to investigate the influence of the occlusal splint on three-element implant-supported fixed prosthesis. Materials and Methods: A three-dimensional virtual model was created consisting of a cortical and spongy bone block simulating the region from first premolar to the maxillary first molar using two HE or MT implants (4 x 11mm) with Ti and/or Y-TZP abutments. The second premolar was the pontic of the prosthesis. The three-element fixed prosthesis with a zirconia infrastructure and Y-TZP coating were cemented, in addition to using a flat occlusal splint made of acrylic resin in the region. Combined axial and oblique loads of 100N and 300N were applied. Results: The tensile stresses on MT implant bone tissue produced values of 4-19% lower than those of HE implants. The lowest differences were observed for oblique loading with an occlusal splint, with a 4% (Ti-Y-TZP) and 9% (Ti-Ti) decrease. When the compressive stresses were evaluated, HE implants produced lower values than MT implants. Conclusion: A significant increase was observed in the oblique loading stresses in the absence of occlusal splints, regardless of the applied load. (Copyright© 2021 Dennis Barber Ltd.) |
فهرسة مساهمة: | Keywords: Finite Element Analysis; Implant Prosthesis; Occlusal Splints; Stress Concentration |
المشرفين على المادة: | 0 (Dental Implants) |
تواريخ الأحداث: | Date Created: 20201020 Date Completed: 20210603 Latest Revision: 20210603 |
رمز التحديث: | 20231215 |
DOI: | 10.1922/EJPRD_02080Baroudi09 |
PMID: | 33079499 |
قاعدة البيانات: | MEDLINE |
تدمد: | 0965-7452 |
---|---|
DOI: | 10.1922/EJPRD_02080Baroudi09 |