دورية أكاديمية

Use of a Checklist Approach on a Telecobalt in an Attempt to Reduce Human Errors in Radiotherapy Delivery and Improve Therapeutic Ratio.

التفاصيل البيبلوغرافية
العنوان: Use of a Checklist Approach on a Telecobalt in an Attempt to Reduce Human Errors in Radiotherapy Delivery and Improve Therapeutic Ratio.
المؤلفون: Mukundan H; Department of Radiation Oncology, Command Hospital (AF), Bengaluru, Karnataka, India., Singh S; Department of Radiation Oncology, Command Hospital (CC), Lucknow, Uttar Pradesh, India., Lohia N; Department of Radiation Oncology, Command Hospital (CC), Lucknow, Uttar Pradesh, India., Semwal MK; Department of Radiation Oncology, Army Hospital (R&R), New Delhi, India., Kumar V; Department of Radiation Oncology, Command Hospital (AF), Bengaluru, Karnataka, India., Bhatnagar S; Department of Radiation Oncology, Army Hospital (R&R), New Delhi, India., Sarin A; Department of Radiation Oncology, Indian Naval Hospital Ship Asvini, Mumbai, Maharashtra, India., Trivedi G; Department of Radiation Oncology, Command Hospital (CC), Lucknow, Uttar Pradesh, India.
المصدر: Journal of medical physics [J Med Phys] 2021 Jan-Mar; Vol. 46 (1), pp. 1-6. Date of Electronic Publication: 2021 May 05.
نوع المنشور: Journal Article
اللغة: English
بيانات الدورية: Publisher: Medknow Publications Country of Publication: India NLM ID: 9441104 Publication Model: Print-Electronic Cited Medium: Print ISSN: 0971-6203 (Print) Linking ISSN: 09716203 NLM ISO Abbreviation: J Med Phys Subsets: PubMed not MEDLINE
أسماء مطبوعة: Publication: Mumbai : Medknow Publications
Original Publication: Bombay : The Association, [1994-
مستخلص: Background: The process of radiotherapy treatment planning and delivery involves multiple steps and professionals causing it to be prone to errors. Radiotherapy centers equipped with old telecobalt machines have certain peculiar challenges to workflow. We designed and tested a checklist for radiotherapy technicians (RTTs) to reduce chances of error during treatment delivery on a telecobalt machine.
Materials and Methods: A physical checklist was designed for RTTs to use in the pretreatment pause using a template advocated by the American Association of Physicists in Medicine. It was tested on 4 RTTs over 1000 radiotherapy delivery sittings.
Results: The checklist helped to rectify 41 documentary lapses and 28 errors in radiotherapy treatment parameters while also identifying 12 instances where treatment plan modifications were due and 30 where the patient was due for review by the radiation oncologist. The average time to go through the checklist was between 2.5 and 3 min.
Conclusions: The development and use of the checklist has helped in reducing errors and also improving workflow in our department. It is recommended to utilize such physical checklists in all radiotherapy centers with telecobalt machines. The success of the checklist depends upon leadership, teamwork, acceptance of a need to inculcate a "safety culture," with voluntary error-reporting and a willingness to learn from such errors.
Competing Interests: There are no conflicts of interest.
(Copyright: © 2021 Journal of Medical Physics.)
References: N Engl J Med. 2006 Dec 28;355(26):2725-32. (PMID: 17192537)
N Engl J Med. 2009 Jan 29;360(5):491-9. (PMID: 19144931)
Radiother Oncol. 2003 Oct;69(1):91-7. (PMID: 14597361)
J Crit Care. 2006 Sep;21(3):231-5. (PMID: 16990087)
Indian J Cancer. 2019 Oct-Dec;56(4):359-363. (PMID: 31607709)
Pract Radiat Oncol. 2018 Jul - Aug;8(4):279-286. (PMID: 29429922)
J Oncol Pract. 2011 Jul;7(4):222-6. (PMID: 22043184)
Clin Oncol (R Coll Radiol). 2019 Jan;31(1):e67-e74. (PMID: 30322681)
J Appl Clin Med Phys. 2015 May 08;16(3):5431. (PMID: 26103502)
Radiother Oncol. 2009 Jul;92(1):15-21. (PMID: 19395105)
J Med Phys. 2014 Oct;39(4):203-5. (PMID: 25525306)
فهرسة مساهمة: Keywords: Checklist; error; radiotherapy; telecobalt
تواريخ الأحداث: Date Created: 20210716 Latest Revision: 20220424
رمز التحديث: 20240628
مُعرف محوري في PubMed: PMC8240907
DOI: 10.4103/jmp.JMP_106_20
PMID: 34267483
قاعدة البيانات: MEDLINE
الوصف
تدمد:0971-6203
DOI:10.4103/jmp.JMP_106_20