دورية أكاديمية

Load-induced deformation of the tibia and its effect on implant loosening detection.

التفاصيل البيبلوغرافية
العنوان: Load-induced deformation of the tibia and its effect on implant loosening detection.
المؤلفون: Ter Wee MA; Department of Biomedical Engineering and Physics, Amsterdam UMC Location University of Amsterdam, Meibergdreef 9, Amsterdam, The Netherlands. m.a.terwee@amsterdamumc.nl.; Amsterdam Movement Sciences, Musculoskeletal Health, Amsterdam, The Netherlands. m.a.terwee@amsterdamumc.nl., Dobbe JGG; Department of Biomedical Engineering and Physics, Amsterdam UMC Location University of Amsterdam, Meibergdreef 9, Amsterdam, The Netherlands.; Amsterdam Movement Sciences, Musculoskeletal Health, Amsterdam, The Netherlands., Buijs GS; Amsterdam Movement Sciences, Musculoskeletal Health, Amsterdam, The Netherlands.; Department of Orthopedic Surgery and Sports Medicine, Amsterdam UMC Location University of Amsterdam, Meibergdreef 9, Amsterdam, The Netherlands., Kievit AJ; Amsterdam Movement Sciences, Musculoskeletal Health, Amsterdam, The Netherlands.; Department of Orthopedic Surgery and Sports Medicine, Amsterdam UMC Location University of Amsterdam, Meibergdreef 9, Amsterdam, The Netherlands., Schafroth MU; Amsterdam Movement Sciences, Musculoskeletal Health, Amsterdam, The Netherlands.; Department of Orthopedic Surgery and Sports Medicine, Amsterdam UMC Location University of Amsterdam, Meibergdreef 9, Amsterdam, The Netherlands., Maas M; Amsterdam Movement Sciences, Musculoskeletal Health, Amsterdam, The Netherlands.; Department of Radiology, Amsterdam UMC Location University of Amsterdam, Meibergdreef 9, Amsterdam, The Netherlands., Blankevoort L; Amsterdam Movement Sciences, Musculoskeletal Health, Amsterdam, The Netherlands.; Department of Orthopedic Surgery and Sports Medicine, Amsterdam UMC Location University of Amsterdam, Meibergdreef 9, Amsterdam, The Netherlands., Streekstra GJ; Department of Biomedical Engineering and Physics, Amsterdam UMC Location University of Amsterdam, Meibergdreef 9, Amsterdam, The Netherlands.; Amsterdam Movement Sciences, Musculoskeletal Health, Amsterdam, The Netherlands.; Department of Radiology, Amsterdam UMC Location University of Amsterdam, Meibergdreef 9, Amsterdam, The Netherlands.
المصدر: Scientific reports [Sci Rep] 2023 Dec 08; Vol. 13 (1), pp. 21769. Date of Electronic Publication: 2023 Dec 08.
نوع المنشور: Journal Article
اللغة: English
بيانات الدورية: Publisher: Nature Publishing Group Country of Publication: England NLM ID: 101563288 Publication Model: Electronic Cited Medium: Internet ISSN: 2045-2322 (Electronic) Linking ISSN: 20452322 NLM ISO Abbreviation: Sci Rep Subsets: MEDLINE
أسماء مطبوعة: Original Publication: London : Nature Publishing Group, copyright 2011-
مواضيع طبية MeSH: Tibia*/diagnostic imaging , Tibia*/surgery , Arthroplasty, Replacement, Knee*/methods, Bone Cements ; Prostheses and Implants ; Image Processing, Computer-Assisted ; Knee Joint/diagnostic imaging ; Knee Joint/surgery
مستخلص: CT imaging under external valgus and varus loading conditions and consecutive image analysis can be used to detect tibial implant loosening after total knee arthroplasty. However, the applied load causes the tibia to deform, which could result in an overestimation of implant displacement. This research evaluates the extent of tibia deformation and its effect on measuring implant displacement. Ten cadaver specimen with TKA were CT-scanned under valgus/varus loading (20 Nm), first implanted without bone cement fixation (mimicking a loose implant) and subsequently with bone cement fixation (mimicking a fixed implant). By means of image analysis, three relative displacements were assessed: (1) between the proximal and distal tibia (measure of deformation), (2) between the implant and the whole tibia (including potential deformation effect) and (3) between the implant and the proximal tibia (reduced deformation effect). Relative displacements were quantified in terms of translations along, and rotations about the axes of a local coordinate system. As a measure of deformation, the proximal tibia moved relative to the distal tibia by, on average 1.27 mm (± 0.50 mm) and 0.64° (± 0.25°). Deformation caused an overestimation of implant displacement in the cemented implant. The implant displaced with respect to the whole tibia by 0.45 mm (± 0.22 mm) and 0.79° (± 0.38°). Relative to the proximal tibia, the implant moved by 0.23 mm (± 0.10 mm) and 0.62° (± 0.34°). The differentiation between loose and fixed implants improved when tibia deformation was compensated for by using the proximal tibia rather than the whole tibia.
(© 2023. The Author(s).)
References: J Bone Joint Surg Br. 2008 Mar;90(3):265-71. (PMID: 18310744)
J Comput Assist Tomogr. 2016 Nov/Dec;40(6):971-978. (PMID: 27331924)
Bone Joint J. 2018 Feb;100-B(2):170-175. (PMID: 29437058)
Acta Orthop. 2021 Aug;92(4):419-423. (PMID: 33821746)
J Biomech. 2002 Apr;35(4):543-8. (PMID: 11934426)
Acta Radiol. 2019 Dec;60(12):1643-1652. (PMID: 31042065)
J Bone Joint Surg Br. 1995 May;77(3):377-83. (PMID: 7744919)
IEEE Trans Med Imaging. 2019 Apr;38(4):1048-1057. (PMID: 30369440)
Osteoporos Int. 2009 Jan;20(1):61-70. (PMID: 18496638)
Clin Biomech (Bristol, Avon). 2023 Apr;104:105930. (PMID: 36906985)
J Clin Densitom. 2006 Apr-Jun;9(2):175-83. (PMID: 16785078)
Knee. 2005 Aug;12(4):301-6. (PMID: 15982891)
J Orthop Res. 2013 Nov;31(11):1686-93. (PMID: 24038514)
IEEE Trans Biomed Eng. 2011 Jan;58(1):182-90. (PMID: 20934945)
ScientificWorldJournal. 2021 Jun 2;2021:8836687. (PMID: 34121950)
Clin Orthop Relat Res. 2005 May;(434):183-8. (PMID: 15864050)
Biomed Res Int. 2016;2016:5909741. (PMID: 27478832)
IEEE Trans Med Imaging. 2001 Sep;20(9):917-27. (PMID: 11585208)
Knee Surg Sports Traumatol Arthrosc. 2011 Sep;19(9):1442-52. (PMID: 21822665)
J Bone Joint Surg Br. 2011 Mar;93(3):293-7. (PMID: 21357948)
J Orthop Res. 2023 Jul 13;:. (PMID: 37442642)
Surg Radiol Anat. 2022 Jun;44(6):851-860. (PMID: 35534775)
Radiographics. 2004 Nov-Dec;24(6):1679-91. (PMID: 15537976)
Acta Radiol. 2020 Jun;61(6):776-782. (PMID: 31684750)
Acta Orthop. 2022 Oct 31;93:831-836. (PMID: 36314542)
Skeletal Radiol. 2019 Oct;48(10):1565-1572. (PMID: 30976835)
Calcif Tissue Res. 1975;17(2):165-8. (PMID: 1139367)
Acta Orthop. 2023 Apr 27;94:207-214. (PMID: 37114404)
J Exp Biol. 1999 Mar;202(Pt 5):543-52. (PMID: 9929457)
Acta Orthop Scand. 2002 Dec;73(6):630-7. (PMID: 12553509)
Acta Orthop. 2020 Dec;91(6):654-659. (PMID: 33063555)
PLoS One. 2014 Apr 14;9(4):e94525. (PMID: 24732724)
Biomed Tech (Berl). 2005 May;50(5):143-7. (PMID: 15966619)
معلومات مُعتمدة: 18547 Nederlandse Organisatie voor Wetenschappelijk Onderzoek
المشرفين على المادة: 0 (Bone Cements)
تواريخ الأحداث: Date Created: 20231208 Date Completed: 20231216 Latest Revision: 20231222
رمز التحديث: 20231222
مُعرف محوري في PubMed: PMC10709436
DOI: 10.1038/s41598-023-49177-z
PMID: 38066256
قاعدة البيانات: MEDLINE
الوصف
تدمد:2045-2322
DOI:10.1038/s41598-023-49177-z