Falsely decreased Abbott Alinity-c gamma-glutamyl transferase-2 result from paraprotein and heparin interference: case report and subsequent laboratory experiments.

التفاصيل البيبلوغرافية
العنوان: Falsely decreased Abbott Alinity-c gamma-glutamyl transferase-2 result from paraprotein and heparin interference: case report and subsequent laboratory experiments.
المؤلفون: Sam KM; Alfred Pathology Service, 5392 Alfred Health , Melbourne, VIC, Australia.; Monash University, Melbourne, VIC, Australia., Schneider HG; Alfred Pathology Service, 5392 Alfred Health , Melbourne, VIC, Australia.; Monash University, Melbourne, VIC, Australia., Choy KW; Department of Pathology, 3234 Northern Hospital Epping , Epping, VIC, Australia., Lam QT; Alfred Pathology Service, 5392 Alfred Health , Melbourne, VIC, Australia.
المصدر: Clinical chemistry and laboratory medicine [Clin Chem Lab Med] 2024 Mar 05; Vol. 62 (6), pp. e126-e128. Date of Electronic Publication: 2024 Mar 05 (Print Publication: 2024).
نوع المنشور: Case Reports; Letter
اللغة: English
بيانات الدورية: Publisher: Walter De Gruyter Country of Publication: Germany NLM ID: 9806306 Publication Model: Electronic-Print Cited Medium: Internet ISSN: 1437-4331 (Electronic) Linking ISSN: 14346621 NLM ISO Abbreviation: Clin Chem Lab Med Subsets: MEDLINE
أسماء مطبوعة: Original Publication: Berlin ; New York : Walter De Gruyter, c1998-
مواضيع طبية MeSH: Heparin* , gamma-Glutamyltransferase*/blood , Paraproteins*/analysis, Humans ; Male ; Female ; False Positive Reactions
References: Alberti, MO, Drake, TA, Song, L. The pH of chemistry assays plays an important role in monoclonal immunoglobulin interferences. Pract Lab Med 2015;3:8–16. https://doi.org/10.1016/j.plabm.2015.09.001 . (PMID: 10.1016/j.plabm.2015.09.001)
Dimeski, G, Carter, A. Rare IgM interference with Roche/Hitachi Modular glucose and gamma-glutamyltransferase methods in heparin samples. Clin Chem 2005;51:2202. https://doi.org/10.1373/clinchem.2005.053561 . (PMID: 10.1373/clinchem.2005.053561)
Graham, L, Dinesh, T, Syme, NR. Prevalence of paraprotein interference in three clinical chemistry analytes in a routine biochemistry laboratory. Ann Clin Biochem 2021;58:667–8. https://doi.org/10.1177/00045632211013862 . (PMID: 10.1177/00045632211013862)
Australian medicines handbook online [Internet] . Protamine [updated Jul 2023]. Australia: Australian Medicines HandbookPty Ltd; 2023. https://amhonline.amh.net.au.acs.hcn.com.au/chapters/blood-electrolytes/drugs-reversing-anticoagulation/protamine [Accessed 12 Sep 2023].
Mylon, E, Winternitz, MC, de Suto-Nagy, GJ. The determination of fibrinogen with protamine. J Biol Chem 1942;143:21–7. https://doi.org/10.1016/S0021-9258(18)72654-9 . (PMID: 10.1016/S0021-9258(18)72654-9)
Strømme, JH, Theodorsen, L. Heparin interference in the measurement of γ-glutamyltransferase activity with the Scandinavian and the IFCC recommended method. Scand J Clin Lab Investig 1985;45:437–42. https://doi.org/10.3109/00365518509155240 . (PMID: 10.3109/00365518509155240)
Gamma-glutamyl transferase2 [package insert] . Longford, Ireland: Abbott Laboratories; 2021.
ɣ-Glutamyltransferase ver.2 – standardized against IFCC/Szasz [package insert] . Mannheim, Germany: Roche Diagnostics GmbH; 2021.
MAbTrap Kit Instructions for Use [package insert]. ©2020 Cytiva. Available from: https://www.cytivalifesciences.com/en/us/shop/mabtrap-kit-p-04814#related-documents .
فهرسة مساهمة: Keywords: GGT-2; gamma-glutamyltransferase (GGT); paraprotein interference
المشرفين على المادة: 9005-49-6 (Heparin)
EC 2.3.2.2 (gamma-Glutamyltransferase)
0 (Paraproteins)
تواريخ الأحداث: Date Created: 20240305 Date Completed: 20240429 Latest Revision: 20240429
رمز التحديث: 20240429
DOI: 10.1515/cclm-2024-0015
PMID: 38439680
قاعدة البيانات: MEDLINE
الوصف
تدمد:1437-4331
DOI:10.1515/cclm-2024-0015