Primary Stability and Fracture Resistance of Zirconia and Titanium Implants: A Paired Comparative In Vitro Study.

التفاصيل البيبلوغرافية
العنوان: Primary Stability and Fracture Resistance of Zirconia and Titanium Implants: A Paired Comparative In Vitro Study.
المؤلفون: Francescato O, Souza Rodrigues IN, Douglas de Oliveira DW, Moreira AN, Magalhães CS, Lages FS
المصدر: The International journal of oral & maxillofacial implants [Int J Oral Maxillofac Implants] 2024 May 31; Vol. 0 (0), pp. 1-22. Date of Electronic Publication: 2024 May 31.
Publication Model: Ahead of Print
نوع المنشور: Journal Article
اللغة: English
بيانات الدورية: Publisher: Quintessence Pub. Co Country of Publication: United States NLM ID: 8611905 Publication Model: Print-Electronic Cited Medium: Internet ISSN: 1942-4434 (Electronic) Linking ISSN: 08822786 NLM ISO Abbreviation: Int J Oral Maxillofac Implants Subsets: MEDLINE
أسماء مطبوعة: Original Publication: [Lombard, Ill. : Quintessence Pub. Co., c1986-
مستخلص: Purpose: Compare Neodent Zi® zirconia implants' insertion torque (IT) and removal torque (RT) with Neodent Alvim® titanium. Measure the maximum torque supported by the zirconia implant until its fracture (MT) and the maximum torque sustained by the assembler of this implant (MTA) until its fracture.
Material and Methods: In this in-vitro study twenty four implants were used. Two groups of implants with the same macrogeometry and from the same manufacturer were compared, Zirconia (n=12) and Titanium (n=12). Implant bed preparations were completed in bovine ribs following a standardized drilling protocol. Then, the insertion torque (IT), removal torque (RT), maximum torque to fracture (MT) and maximum torque to fracture of the assembler (MTA) were completed using a calibrated torque meter. Data was presented using descriptive statistics including means, standard deviations (SD), medians, and quartiles. The Shapiro-Wilk test was used to verify data normality and the Wilcoxon test was used to evaluate differences between groups. Statistical significance was established as p < 0.05.
Results: Zirconia implants showed: IT 89.33 ± 31.18 Ncm and RT 84.89 ± 32.92 Ncm. Titanium implants showed: IT 77.58 ± 28.96 Ncm and RT 76.75 ± 31.29 Ncm without significant differences (p>0.05). In relation to fracture under rotational force, the zirconia implants fractured at 106.17 ± 22.54 Ncm, and the implant assembly fractured at 84.00 ±13.14 Ncm.
Conclusion: Neodent Zi® implants showed stability but lower fracture torque than Alvim® titanium. As the fracture values of the assembler were significantly lower than the fracture values of the zirconia implants, it can be stated that, as recommended by the manufacturer, they act as a safety measure during installation.