دورية أكاديمية

Long-Term Effects of Pulmonary Valve Implantation and Prosthesis Evolution in Patients with Repaired Tetralogy of Fallot.

التفاصيل البيبلوغرافية
العنوان: Long-Term Effects of Pulmonary Valve Implantation and Prosthesis Evolution in Patients with Repaired Tetralogy of Fallot.
عنوان ترانسليتريتد: Efeitos de Longo Prazo do Implante Valvar Pulmonar e Evolução da Prótese em Pacientes com Tetralogia de Fallot Corrigida.
المؤلفون: Caneo LF; Instituto do Coração do Hospital das Clínicas da Faculdade de Medicina da Universidade de São Paulo, São Paulo, SP - Brasil., Turquetto ALR; Instituto do Coração do Hospital das Clínicas da Faculdade de Medicina da Universidade de São Paulo, São Paulo, SP - Brasil., Boschiero MN; Universidade São Francisco Faculdade de Medicina, Bragança Paulista, SP - Brasil., Amato LP; Instituto do Coração do Hospital das Clínicas da Faculdade de Medicina da Universidade de São Paulo, São Paulo, SP - Brasil., Ishikawa WY; Instituto do Coração do Hospital das Clínicas da Faculdade de Medicina da Universidade de São Paulo, São Paulo, SP - Brasil., Hodas FP; Instituto do Coração do Hospital das Clínicas da Faculdade de Medicina da Universidade de São Paulo, São Paulo, SP - Brasil., Ligeiro MG; Instituto do Coração do Hospital das Clínicas da Faculdade de Medicina da Universidade de São Paulo, São Paulo, SP - Brasil., Agostinho DR; Instituto do Coração do Hospital das Clínicas da Faculdade de Medicina da Universidade de São Paulo, São Paulo, SP - Brasil., Miana LA; Instituto do Coração do Hospital das Clínicas da Faculdade de Medicina da Universidade de São Paulo, São Paulo, SP - Brasil., Tanamati C; Instituto do Coração do Hospital das Clínicas da Faculdade de Medicina da Universidade de São Paulo, São Paulo, SP - Brasil., Gonçalves RC; Instituto do Coração do Hospital das Clínicas da Faculdade de Medicina da Universidade de São Paulo, São Paulo, SP - Brasil., Penha JG; Instituto do Coração do Hospital das Clínicas da Faculdade de Medicina da Universidade de São Paulo, São Paulo, SP - Brasil., Massoti MRB; Instituto do Coração do Hospital das Clínicas da Faculdade de Medicina da Universidade de São Paulo, São Paulo, SP - Brasil., Jatene MB; Instituto do Coração do Hospital das Clínicas da Faculdade de Medicina da Universidade de São Paulo, São Paulo, SP - Brasil., Jatene FB; Instituto do Coração do Hospital das Clínicas da Faculdade de Medicina da Universidade de São Paulo, São Paulo, SP - Brasil.
المصدر: Arquivos brasileiros de cardiologia [Arq Bras Cardiol] 2024 Jun; Vol. 121 (7), pp. e20230585.
نوع المنشور: Journal Article
اللغة: Portuguese; English
بيانات الدورية: Publisher: Sociedad Brasileira De Cardiologia Country of Publication: Brazil NLM ID: 0421031 Publication Model: Print Cited Medium: Internet ISSN: 1678-4170 (Electronic) Linking ISSN: 0066782X NLM ISO Abbreviation: Arq Bras Cardiol Subsets: MEDLINE
أسماء مطبوعة: Original Publication: Sao Paulo : Sociedad Brasileira De Cardiologia
مواضيع طبية MeSH: Tetralogy of Fallot*/surgery , Tetralogy of Fallot*/physiopathology , Tetralogy of Fallot*/diagnostic imaging , Pulmonary Valve Insufficiency*/surgery , Pulmonary Valve Insufficiency*/physiopathology , Pulmonary Valve Insufficiency*/diagnostic imaging , Pulmonary Valve Insufficiency*/etiology , Heart Valve Prosthesis Implantation* , Pulmonary Valve*/surgery , Pulmonary Valve*/diagnostic imaging , Pulmonary Valve*/physiopathology , Heart Valve Prosthesis*, Humans ; Male ; Female ; Retrospective Studies ; Treatment Outcome ; Adult ; Adolescent ; Young Adult ; Time Factors ; Child ; Echocardiography ; Ventricular Function, Right/physiology ; Stroke Volume/physiology ; Heart Ventricles/physiopathology ; Heart Ventricles/diagnostic imaging ; Magnetic Resonance Imaging ; Postoperative Period
مستخلص: Background: Pulmonary valve regurgitation is a significant long-term complication in patients with tetralogy of Fallot (TOF).
Objective: This study aims to investigate the effects of pulmonary valve implantation (PVI) on the anatomy and function of the right ventricle (RV) and the long-term evolution of the implanted prosthesis in the pulmonary position.
Methods: A single-center retrospective cohort analysis was performed in 56 consecutive patients with TOF who underwent PVI. The study included patients of both sexes, aged ≥ 12 years, and involved assessing clinical and surgical data, pre- and post-operative cardiovascular magnetic resonance imaging, and echocardiogram data more than 1 year after PVI.
Results: After PVI, there was a significant decrease in RV end-systolic volume indexed by body surface area (BSA), from 89 mL/BSA to 69 mL/BSA (p < 0.001) and indexed RV end-diastolic volume, from 157 mL/BSA to 116 mL/BSA (p < 0.001). Moreover, there was an increase in corrected RV ejection fraction [ RVEFC = net pulmonary flow (pulmonary forward flow - regurgitant flow) / R V end-diastolic volume ] from 23% to 35% (p < 0.001) and left ventricular ejection fraction from 58% to 60% (p = 0.008). However, a progressive increase in the peak pulmonary valve gradient was observed over time, with 25% of patients experiencing a gradient exceeding 60 mmHg. Smaller prostheses (sizes 19 to 23) were associated with a 4.3-fold higher risk of a gradient > 60 mmHg compared to larger prostheses (sizes 25 to 27; p = 0.029; confidence interval: 1.18 to 17.8).
Conclusion: As expected, PVI demonstrated improvements in RV volumes and function. Long-term follow-up and surveillance are crucial for assessing the durability of the prosthesis and detecting potential complications. Proper sizing of prostheses is essential for improved prosthesis longevity.
فهرسة مساهمة: Local Abstract: [Publisher, Portuguese] Figura Central : Efeitos de Longo Prazo do Implante Valvar Pulmonar e Evolução da Prótese em Pacientes com Tetralogia de Fallot Corrigida TF: tetralogia de Fallot. [Publisher, Portuguese] A regurgitação valvar pulmonar é uma importante complicação de longo prazo em pacientes com tetralogia de Fallot (TF). [Publisher, Portuguese] O presente estudo tem como objetivo investigar os efeitos do implante valvar pulmonar (IVP) na anatomia e função do ventrículo direito (VD) e na evolução em longo prazo da prótese implantada em posição pulmonar. [Publisher, Portuguese] Uma análise de coorte retrospectiva e unicêntrica foi realizada em 56 pacientes consecutivos com TF submetidos a IVP. O estudo incluiu pacientes de ambos os gêneros, com idade ≥ 12 anos e compreendeu avaliação de dados clínicos e cirúrgicos, ressonância magnética cardiovascular pré e pós-operatória e dados ecocardiográficos obtidos mais de 1 ano após IVP. [Publisher, Portuguese] Após o IVP, houve uma diminuição significativa do volume sistólico final do VD indexado pela área de superfície corpórea (ASC), de 89 mL/ASC para 69 mL/ASC (p < 0,001) e do volume diastólico final indexado do VD, de 157 mL/ASC para 116 mL/ASC (p < 0,001). Além disso, houve aumento da fração de ejeção corrigida do VD [ FEVDc = fluxo pulmonar ajustado (fluxo pulmonar anterógrado − fluxo regurgitante) / volume diastólico final do VD ] de 23% para 35% (p < 0,001) e da fração de ejeção do ventrículo esquerdo de 58% para 60% (p = 0,008). No entanto, foi observado um aumento progressivo no gradiente de pico da válvula pulmonar ao longo do tempo, com 25% dos pacientes apresentando um gradiente superior a 60 mmHg. Próteses menores (tamanhos 19 a 23) foram associadas a um risco 4,3 vezes maior de gradiente > 60 mmHg em comparação com próteses maiores (tamanhos 25 a 27; p = 0,029; intervalo de confiança: 1,18 a 17,8). [Publisher, Portuguese] Conforme esperado, o IVP demonstrou melhorias nos volumes e na função do VD. O acompanhamento e a vigilância a longo prazo são cruciais para avaliar a durabilidade da prótese e detectar potenciais complicações. O dimensionamento adequado das próteses é essencial para melhorar a longevidade da prótese.
تواريخ الأحداث: Date Created: 20240731 Date Completed: 20240731 Latest Revision: 20240731
رمز التحديث: 20240731
DOI: 10.36660/abc.20230585
PMID: 39082573
قاعدة البيانات: MEDLINE
الوصف
تدمد:1678-4170
DOI:10.36660/abc.20230585