دورية أكاديمية

Ionization quenching in scintillators used for dosimetry of mixed particle fields.

التفاصيل البيبلوغرافية
العنوان: Ionization quenching in scintillators used for dosimetry of mixed particle fields.
المؤلفون: Jeppe Brage Christensen, Erik Almhagen, Liliana Stolarczyk, Anne Vestergaard, Niels Bassler, Claus E Andersen
المصدر: Physics in Medicine & Biology; May2019, Vol. 64 Issue 9, p1-1, 1p
مصطلحات موضوعية: SCINTILLATORS, LINEAR energy transfer, PARTICLE tracks (Nuclear physics), RADIATION dosimetry, PHOTON beams, ION beams, PROTONS
مستخلص: Ionization quenching in ion beam dosimetry is often related to the fluence- or dose-averaged linear energy transfer (LET). Both quantities are however averaged over a wide LET range and a mixed field of primary and secondary ions. We propose a novel method to correct the quenched luminescence in scintillators exposed to ion beams. The method uses the energy spectrum of the primaries and accounts for the varying quenched luminescence in heavy, secondary ion tracks through amorphous track structure theory. The new method is assessed against more traditional approaches by correcting the quenched luminescence response from the BCF-12, BCF-60, and 81-0084 plastic scintillators exposed to a 100 MeV pristine proton beam in order to compare the effects of the averaged LET quantities and the secondary ions. Calculations and measurements show that primary protons constitute more than 92% of the energy deposition but account for more than 95% of the luminescence signal in the scintillators. The quenching corrected luminescence signal is in better agreement with the dose measurement when the secondary particles are taken into account. The Birks model provided the overall best quenching corrections, when the quenching corrected signal is adjusted for the number of free model parameters. The quenching parameter kB for the BCF-12 and BCF-60 scintillators is in agreement with literature values and was found to be m keV−1 for the 81-0084 scintillator. Finally, a fluence threshold for the 100 MeV proton beam was calculated to be of the order of 1010 cm−2, corresponding to 110 Gy, above which the quenching increases non-linearly and the Birks model no longer is applicable. [ABSTRACT FROM AUTHOR]
Copyright of Physics in Medicine & Biology is the property of IOP Publishing and its content may not be copied or emailed to multiple sites or posted to a listserv without the copyright holder's express written permission. However, users may print, download, or email articles for individual use. This abstract may be abridged. No warranty is given about the accuracy of the copy. Users should refer to the original published version of the material for the full abstract. (Copyright applies to all Abstracts.)
قاعدة البيانات: Complementary Index
الوصف
تدمد:00319155
DOI:10.1088/1361-6560/ab12f2