دورية أكاديمية

Retrieval analysis of neck fracture on uni-modular total hip arthroplasty stems: The contributions of material processing and stem design.

التفاصيل البيبلوغرافية
العنوان: Retrieval analysis of neck fracture on uni-modular total hip arthroplasty stems: The contributions of material processing and stem design.
المؤلفون: Cubillos, Patricia O, dos Santos, Vinícius O, Fernandes, Daniel A, Moré, Ari DO, da Rosa, Edison, Fancello, Eduardo A, Roesler, Carlos RM
المصدر: Proceedings of the Institution of Mechanical Engineers -- Part H -- Journal of Engineering in Medicine (Sage Publications, Ltd.); Feb2022, Vol. 236 Issue 2, p147-157, 11p
مستخلص: Total hip arthroplasty stem fracture is an important contributor to morbidity rate and increases the cost of revision surgery. Failure is usually caused by issues related to overload, inadequate stem support, inappropriate stem design or dimensions and material processing. In this study, the role of the relationship between material characterization and biomechanical performance in the fracture of retrieved stems was explored. The stems were manufactured with forged stainless steel, had the same length, 12/14 trunnion, and 28-mm head. These stems were evaluated by macroscopic and microscopic examination to identify the causes of premature failure. Each stem was sectioned into four regions, and the cross-sections were used for the microhardness and grain size analysis. Finite element analysis (FEA) was carried out, considering the stem positioned at the femur, a musculoskeletal model, and biomechanical loading. All stems had fractured through a fatigue mechanism, mainly a unidirectional bending loading condition, with crack nucleation on the lateral side and propagation on the medial side. The numerical analysis revealed maximum mechanical stress on the lateral side of the stem neck, but this was below the yield stress calculated via the hardness. The use of a shorter head neck length could reduce the maximum mechanical stress at the neck. At a cross-section near the plane of the stem fracture, the hardness was lower than that normally reported by the ASM, and there were heterogonous and coarse grain sizes on the lateral side. The main cause of failure of the two stems analyzed was a combination of low hardness and coarse grain size, due to inappropriate materials processing, worsen by a high level of stress on the lateral side of the neck due to the large stem-head offset selected by the orthopedic surgeon. [ABSTRACT FROM AUTHOR]
Copyright of Proceedings of the Institution of Mechanical Engineers -- Part H -- Journal of Engineering in Medicine (Sage Publications, Ltd.) is the property of Sage Publications, Ltd. and its content may not be copied or emailed to multiple sites or posted to a listserv without the copyright holder's express written permission. However, users may print, download, or email articles for individual use. This abstract may be abridged. No warranty is given about the accuracy of the copy. Users should refer to the original published version of the material for the full abstract. (Copyright applies to all Abstracts.)
قاعدة البيانات: Complementary Index
الوصف
تدمد:09544119