دورية أكاديمية

A Monovalent Fab Affinity-Capture and Elution Bridging Immunoassay Overcomes Rheumatoid Factor Interference while Accurately Detecting Antidrug Antibodies.

التفاصيل البيبلوغرافية
العنوان: A Monovalent Fab Affinity-Capture and Elution Bridging Immunoassay Overcomes Rheumatoid Factor Interference while Accurately Detecting Antidrug Antibodies.
المؤلفون: Calderon, Boris, Reynolds, Nichole A, Swearingen, Craig A, Carter, Quincy L, Bivi, Nicoletta, Cramer, Jeff W, Alexander, Courtney L, Siegel, Robert W, Konrad, Robert J
المصدر: Journal of Applied Laboratory Medicine; Sep2023, Vol. 8 Issue 5, p896-908, 13p
مصطلحات موضوعية: RHEUMATOID factor, IMMUNOASSAY, IMMUNOGLOBULIN G, MONOCLONAL antibodies, IMMUNOGLOBULINS, IMMUNE response, AUTOANTIBODIES
مستخلص: Background: Rheumatoid factor (RF) consists of autoantibodies that bind the fragment crystallizable (Fc) region of human immunoglobulin G (IgG) and present in sera of rheumatoid arthritis (RA) patients. Immunoassays to detect antidrug antibodies (ADA) in RA patient samples may experience interference due to RF binding and crosslinking Fc regions of the capture and detection antibody reagents. To overcome this interference, a novel Fab affinity-capture and elution (ACE)-bridging immunoassay (Fab ACE-Bridge) was developed with monovalent-recombinant Fab to avoid RF interference. Methods: ACE and ACE-Bridge assays were developed to detect ADA against a therapeutic monoclonal antibody using samples from healthy donors, psoriasis patients, and RA patients. The performance of these assays was compared to a novel Fab ACE-Bridge assay, in which monoclonal antibody was replaced with monovalent Fab. Results: High screening signals in the ACE and ACE-Bridge assays were detected in RA patient samples but not in samples from healthy donors or psoriasis patients. The high screening signals in RA samples did not inhibit to the expected extent in the confirmatory assay, a consistent feature of false-positive screening results. Further investigation revealed RF as the interferent affecting assay performance. Modification of the ACE-Bridge assay by using monovalent Fab eliminated RF interference while allowing for sensitive and drug-tolerant detection of authentic ADA. Conclusions: RF interfered significantly in traditional ACE and ACE-Bridge assays. Implementation of a novel monovalent Fab ACE-Bridge assay overcame RF interference. The use of monovalent Fab is recommended for immunogenicity assays when assessing ADA in RA patient samples. [ABSTRACT FROM AUTHOR]
Copyright of Journal of Applied Laboratory Medicine is the property of Oxford University Press / USA and its content may not be copied or emailed to multiple sites or posted to a listserv without the copyright holder's express written permission. However, users may print, download, or email articles for individual use. This abstract may be abridged. No warranty is given about the accuracy of the copy. Users should refer to the original published version of the material for the full abstract. (Copyright applies to all Abstracts.)
قاعدة البيانات: Complementary Index
الوصف
تدمد:24757241
DOI:10.1093/jalm/jfad035