العنوان: Contaminated Duodenoscopes in ERCP: Assessing Risk and Culture Sensitivity.
المؤلفون: van der Ploeg, K., Severin, J. A., Klaassen, C.H. W., Vos, M. C., Bruno, M. J., Mason-Slingerland, B.C.G. C.
المصدر: Endoscopy; 2024 Supplement 2, Vol. 56, pS368-S369, 2p
مصطلحات موضوعية: ENDOSCOPIC retrograde cholangiopancreatography
مستخلص: This article, published in the journal Endoscopy, discusses the risk of contaminated duodenoscopes in Endoscopic Retrograde Cholangiopancreatography (ERCP) procedures. The study analyzes seven years of microbiological surveillance data to estimate the prevalence of contaminated duodenoscope usage and determine the sensitivity of duodenoscope cultures. The results show that between 16.1% and 23.8% of ERCPs during the study period involved contaminated duodenoscopes, and duodenoscope culture sensitivity ranged from 75.1% to 98.9%. The study concludes that despite surveillance and quarantine measures, there is still a significant risk of patient exposure to contaminated duodenoscopes, and lifting quarantine only after multiple negative cultures should be considered. [Extracted from the article]
Copyright of Endoscopy is the property of Georg Thieme Verlag Stuttgart and its content may not be copied or emailed to multiple sites or posted to a listserv without the copyright holder's express written permission. However, users may print, download, or email articles for individual use. This abstract may be abridged. No warranty is given about the accuracy of the copy. Users should refer to the original published version of the material for the full abstract. (Copyright applies to all Abstracts.)
قاعدة البيانات: Complementary Index
الوصف
تدمد:0013726X
DOI:10.1055/s-0044-1783606