Correction of axial length measurement error by IOLMaster 700 could improve refractive prediction accuracy in silicone oil‐filled eyes.

التفاصيل البيبلوغرافية
العنوان: Correction of axial length measurement error by IOLMaster 700 could improve refractive prediction accuracy in silicone oil‐filled eyes.
المؤلفون: Zhang, Jiaqing, Han, Xiaotong, Jin, Aixia, Zhang, Yifan, Chen, Xiaoyun, Liu, Zhenzhen, Qiu, Xiaozhang, Tan, Xuhua, Luo, Lixia, Liu, Yizhi
المصدر: Acta Ophthalmologica (1755375X); Aug2024, Vol. 102 Issue 5, pe718-e726, 9p
مصطلحات موضوعية: MEASUREMENT errors, LENGTH measurement, RADIAL basis functions, SILICONES, INTRAOCULAR lenses
مستخلص: Purpose: To determine whether correcting the axial length (AL) measurement error of the IOLMaster 700 could improve the refractive prediction accuracy in silicone oil‐filled eyes. Methods: This study included 265 cataract patients (265 eyes) with silicone oil tamponade who were scheduled for phacoemulsification with intraocular lens (IOL) implantation. The performances of various formulas, including Barrett Universal II, Emmetropia Verifying Optical, Hoffer‐QST, Kane, Ladas Super Formula, Pearl‐DGS, Radial Basis Function and traditional formulas (Haigis, Hoffer Q, Holladay 1 and SRK/T), were evaluated. The refractive prediction errors (PE) calculated with measured AL (ALmeas) and corrected AL with silicone oil adjustment (SOAL) were compared. Subgroup analysis was performed based on the ALmeas (<23 mm; 23–26 mm; ≥26 mm). Results: Using SOAL significantly reduced the hyperopic PE of formulas when compared to ALmeas (−0.05 to 0.17 D vs 0.15 to 0.38 D, p < 0.001). After applying AL correction, all formulas showed a lower mean absolute PE (0.47–0.57 D vs 0.50–0.69 D). The percentage of eyes within ±1.0 D of PE increased from 84.91%–88.68% to 89.81%–91.32% for new formulas and from 78.11%–83.40% to 85.66%–88.68% for traditional formulas, with the use of SOAL. Subgroup analysis showed that the majority of formulas with SOAL in prediction accuracy for eyes with an AL ≥26 mm (p < 0.05). Conclusions: The refractive prediction accuracy in silicone oil‐filled eyes was improved by correcting the AL measurement error of the IOLMaster 700, especially for long eyes. [ABSTRACT FROM AUTHOR]
Copyright of Acta Ophthalmologica (1755375X) is the property of Wiley-Blackwell and its content may not be copied or emailed to multiple sites or posted to a listserv without the copyright holder's express written permission. However, users may print, download, or email articles for individual use. This abstract may be abridged. No warranty is given about the accuracy of the copy. Users should refer to the original published version of the material for the full abstract. (Copyright applies to all Abstracts.)
قاعدة البيانات: Complementary Index
الوصف
تدمد:1755375X
DOI:10.1111/aos.16627