Noise power spectrum (NPS) characteristics on CT imaging: Slice thickness and filter types variations.

التفاصيل البيبلوغرافية
العنوان: Noise power spectrum (NPS) characteristics on CT imaging: Slice thickness and filter types variations.
المؤلفون: Raharja, Hanendya Disha Randy, Putra, Doharmansyah, Naufal, Ariij, Anam, Choirul, Supit, Nina I. S. H., Haposan, Ucok Noptua
المصدر: AIP Conference Proceedings; 2024, Vol. 3210 Issue 1, p1-7, 7p
مصطلحات موضوعية: COMPUTED tomography, POWER spectra, FOURIER transforms, NOISE, RADIOLOGISTS
مستخلص: Utilization of high-quality computed tomography (CT) imaging holds great potential in assisting clinical radiologists in diagnosing patient's illnesses with great precision and accuracy. One principal parameter of CT imaging quality is noise level, quantified as noise power spectrum (NPS). The study utilized a head water phantom, consisting of homogeneous material with a diameter of 16 cm. In this work, we performed CT imaging quality assessment by analyzing noise power spectrum (NPS) while varying slice thickness and filter types. The NPS was calculated by performing Fourier Transform on five regions of interest (ROI) along the x- and y- axes using IndoQCT. The slice thickness was varied: 0.625, 1.25, 2.5, 5, and 10 mm. It shows that slice thickness 0.625 mm has the finest sharpness but also has highest noise level with NPS peak value of 255.63 HU2mm2 at frequency of 0.23 mm-1. Slice thickness 10 mm has the lowest sharpness but also has lowest noise level of with NPS peak value of 19.79 HU2mm2 at frequency of 0.2 mm-1. Similarly, we varied the filter types, which included smooth, standard, sharp, detail, standard edge, sharp edge, brain smooth, brain standard, and brain sharp. It shows that filter detail has the finest sharpness but also has highest noise level with NPS peak value of 74.6 HU2mm2 at frequency of 0.54 mm-1. Filter brain smooth has the lowest sharpness but also has lowest noise level of with NPS peak value of 15.9 HU2mm2 at frequency of 0.26 mm-1. Results show that there is an inverse proportionality between slice thickness and NPS value at all frequency levels; thicker slice thickness leads to lower NPS value. NPS increases with filters that provide finer image detail. The choice of slice thickness and filter types depends on the object being diagnosed. [ABSTRACT FROM AUTHOR]
Copyright of AIP Conference Proceedings is the property of American Institute of Physics and its content may not be copied or emailed to multiple sites or posted to a listserv without the copyright holder's express written permission. However, users may print, download, or email articles for individual use. This abstract may be abridged. No warranty is given about the accuracy of the copy. Users should refer to the original published version of the material for the full abstract. (Copyright applies to all Abstracts.)
قاعدة البيانات: Complementary Index
الوصف
تدمد:0094243X
DOI:10.1063/5.0228059