In Vitro Evaluation of the Removal Force of Abutments in Frictional Dental Implants.

التفاصيل البيبلوغرافية
العنوان: In Vitro Evaluation of the Removal Force of Abutments in Frictional Dental Implants.
المؤلفون: Zielak, João César, Rorbacker, Murilo, Gomes, Rodrigo, Yamashita, Celso, Gonzaga, Carla Castiglia, Giovanni, Allan Fernando
المصدر: Journal of Oral Implantology; 2011, Vol. 37 Issue 5, p519-523, 6p
مصطلحات موضوعية: DENTAL abutments, DENTAL implants, DENTURES, ARTIFICIAL implants, TISSUE fixation (Histology)
مستخلص: The objective of the present work was to determine some force parameters for removal of an abutment from a dental implant in a frictional system (locking taper, 1.23 degrees). Ten implants of the same length (11 mm) and different diameters were selected, along with 10 straight abutments (13 mm length) with different diameters. Abutments were attached to implants without application of force. Fixation of the implant-abutment mount (IA) (repeated 1-5 times) was performed through the impact of a body weight (compression force, tapping) left from a known height. After each group of tappings, IA mounts were coupled with a tensile strength tester. The lowest removal value was found after the first tapping of mount #2 (83 N, implant diameter 3.3 mm/4.5 mm abutment diameter), and the highest removal value happened with mount #8 after the fifth tapping (420 N, 5.0 mm/5.5 mm). The force to remove IA mounts increased with the number of tappings and with the increase in abutment mass. Three activations (tappings) of the abutment were considered necessary to yield optimal stability, demonstrated by the large increase in removal force. [ABSTRACT FROM AUTHOR]
Copyright of Journal of Oral Implantology is the property of KnowledgeWorks Global, Ltd and its content may not be copied or emailed to multiple sites or posted to a listserv without the copyright holder's express written permission. However, users may print, download, or email articles for individual use. This abstract may be abridged. No warranty is given about the accuracy of the copy. Users should refer to the original published version of the material for the full abstract. (Copyright applies to all Abstracts.)
قاعدة البيانات: Complementary Index
الوصف
تدمد:01606972
DOI:10.1563/AAID-JOI-D-09-00106.1