Accuracy of two devices and three different calculation methods for predicting residual astigmatism after intraocular lens implantation.

التفاصيل البيبلوغرافية
العنوان: Accuracy of two devices and three different calculation methods for predicting residual astigmatism after intraocular lens implantation.
المؤلفون: Bicknell, Kendall, Helpert, Christopher, Dwyer, Kourtney, McKnight, Brett, Singh, Sonali, Fulcher, Samuel, Fung, Derrick
المصدر: Baylor University Medical Center Proceedings; Apr-Jun2020, Vol. 33 Issue 2, p205-208, 4p, 1 Chart
مستخلص: This study examined how different methods of biometric measurement, types of intraocular lens calculation, and levels of surgeon experience affected predicted postoperative astigmatism with toric intraocular lens implantation. The error in predicted residual astigmatism was determined using the IOL Master 500 and Lenstar LS 900 devices and was calculated by the Alcon toric calculator, Alcon toric calculator with Baylor nomogram, and Barrett toric calculator. Results were reviewed from 147 eyes of 125 patients. Of the three calculations, the Barrett toric calculator had the lowest mean absolute error in predicted residual astigmatism (P < 0.0001). The mean absolute errors of the IOL Master 500 and the Lenstar LS 900 were not significantly different when compared using three different calculation methods (P = 0.54). The mean absolute error was significantly lower (P = 0.049) for ophthalmology staff than for ophthalmology residents when calculation methods were considered. In conclusion, the calculation method used for IOL selection in cataract surgery significantly affects the accuracy of both staff and residents. [ABSTRACT FROM AUTHOR]
Copyright of Baylor University Medical Center Proceedings is the property of Taylor & Francis Ltd and its content may not be copied or emailed to multiple sites or posted to a listserv without the copyright holder's express written permission. However, users may print, download, or email articles for individual use. This abstract may be abridged. No warranty is given about the accuracy of the copy. Users should refer to the original published version of the material for the full abstract. (Copyright applies to all Abstracts.)
قاعدة البيانات: Supplemental Index
الوصف
تدمد:08998280
DOI:10.1080/08998280.2020.1724846