Calibration of diffractive micromirror arrays for microscopy applications

التفاصيل البيبلوغرافية
العنوان: Calibration of diffractive micromirror arrays for microscopy applications
المؤلفون: Berndt, Dirk, Heber, Jörg, Sinning, Steffen, Rudloff, Dirk, Wolschke, Steffen, Eckert, Mark, Schmidt, Jan-Uwe, Bring, Martin, Wagner, Michael, Lakner, Hubert
المصدر: Proceedings of SPIE; June 2011, Vol. 8191 Issue: 1 p81910O-81910O-10, 8109101p
مستخلص: We report on our investigation to precisely actuate diffractive micromirror arrays (MMA) with an accuracy of /100. The test samples consist of analog, torsional MEMS arrays with 65 536 (256x256) mirror elements. These light modulators were developed for structured illumination purposes to be applied as programmable mask for life science and semiconductor microscopy application. Main part of the work relies on the well known characterization of MEMS mirrors with profilometry to automatically measure and approximate the MMA actuation state with high resolution. Examples illustrate the potential of this strategy to control the tilt state of many thousand micromirrors within the accuracy range of the characterization tool. In a dynamic range between 0 and >250 nm the MMA deflection has been precisely adjusted for final MMA application in the deep-UV - VIS - NIR spectral range. The optical properties of calibrated MMAs are tested in a laser measurement setup. After MMA calibration an increased homogeneity and improved image contrast are demonstrated for various illumination patterns.
قاعدة البيانات: Supplemental Index
الوصف
تدمد:0277786X
DOI:10.1117/12.899881