Accident hypoxique en anesthésie pédiatrique dû à un réglage inadapté du respirateur

التفاصيل البيبلوغرافية
العنوان: Accident hypoxique en anesthésie pédiatrique dû à un réglage inadapté du respirateur
المؤلفون: K. Samii, Thierry Garnaud
المصدر: Annales Françaises d'Anesthésie et de Réanimation. 32:872-875
بيانات النشر: Elsevier BV, 2013.
سنة النشر: 2013
مصطلحات موضوعية: Insufflation, business.product_category, business.industry, medicine.medical_treatment, Tracheal intubation, General Medicine, Bronchospasm, Anesthesiology and Pain Medicine, High pressure, Anesthesia, medicine, Breathing, Intubation, Respirator, Respiratory system, medicine.symptom, business
الوصف: A 3-year-old child was anesthetized for ENT examination and surgery. After induction and tracheal intubation, the patient was ventilated (controlled mode). The respirator screen showed information compatible with a failure of intubation: no expired CO2, no expired flow, no alarm of high pressure limit, and no respiratory chest movement. A fall of SpO2 appeared rapidly which recovered after extubation and manual ventilation through a face mask and reintubation. The expiratory CO2 was present when the patient was ventilated manually and disappeared under controlled ventilation. The increase in the value of the maximal insufflation pressure allowed efficient ventilation with an expiratory CO2 curve and showed high ventilation pressure compatible with a bronchospasm. This case report shows that in case of bronchospasm, if the value of the maximal insufflation pressure is low, this may lead to an erroneous diagnosis of failure of intubation.
تدمد: 0750-7658
URL الوصول: https://explore.openaire.eu/search/publication?articleId=doi_________::16771015f9145f63bd81a4195520d839
https://doi.org/10.1016/j.annfar.2013.09.012
رقم الأكسشن: edsair.doi...........16771015f9145f63bd81a4195520d839
قاعدة البيانات: OpenAIRE