Development of a mobile radon calibration system at KRISS

التفاصيل البيبلوغرافية
العنوان: Development of a mobile radon calibration system at KRISS
المؤلفون: Byungjoo Kim, M.J. Han, Dong-Hye Heo, B. C. Kim, J.M. Lee, K.B. Lee, Y. G. Seon, Seung-Mi Hwang
المصدر: Journal of Radioanalytical and Nuclear Chemistry. 330:571-576
بيانات النشر: Springer Science and Business Media LLC, 2021.
سنة النشر: 2021
مصطلحات موضوعية: Health, Toxicology and Mutagenesis, Detector, Public Health, Environmental and Occupational Health, chemistry.chemical_element, Radon, Pollution, Analytical Chemistry, Nuclear Energy and Engineering, chemistry, Calibration, Environmental science, Radiology, Nuclear Medicine and imaging, Spectroscopy, Remote sensing
الوصف: A mobile radon calibration system was developed with a high-sensitivity radon detector, a movable calibration chamber, and a 226Ra solid source at the Korea Research Institute of Standards and Science (KRISS). The high-sensitivity radon detector was based on electrostatic collection mechanism. The volume of the calibration chamber was (1003.8 ± 6.0) L. The mobile radon calibration system was calibrated with 222Rn CRM from the radon primary system at KRISS. The sensitivities were established to be (231.8 ± 16.1) cpm/(kBq/m3) for 218Po and (254.9 ± 17.8) cpm/(kBq/m3) for 214Po, respectively (k = 2).
تدمد: 1588-2780
0236-5731
URL الوصول: https://explore.openaire.eu/search/publication?articleId=doi_________::19edc10eff3fd658d5de16a0d3b05e65
https://doi.org/10.1007/s10967-021-07932-4
حقوق: CLOSED
رقم الأكسشن: edsair.doi...........19edc10eff3fd658d5de16a0d3b05e65
قاعدة البيانات: OpenAIRE