An interlaboratory comparison study on the measurement of elements in PM10

التفاصيل البيبلوغرافية
العنوان: An interlaboratory comparison study on the measurement of elements in PM10
المؤلفون: Krystyna Trzepla, Giulia Calzolai, Michel Gerboles, Claudio A. Belis, Fabrizia Cavalli, Franco Lucarelli, Sinan Yatkin
المصدر: Atmospheric Environment. 125:61-68
بيانات النشر: Elsevier BV, 2016.
سنة النشر: 2016
مصطلحات موضوعية: Detection limit, Atmospheric Science, Materials science, 010504 meteorology & atmospheric sciences, Analytical chemistry, Absolute difference, 010501 environmental sciences, Mass spectrometry, 01 natural sciences, Calibration, Comparison study, Total least squares, 0105 earth and related environmental sciences, General Environmental Science
الوصف: An inter-laboratory comparison study was conducted to measure elemental loadings on PM 10 samples, collected in Ispra, a regional background/rural site in Italy, using three different XRF (X-ray Fluorescence) methods, namely Epsilon 5 by linear calibration, Quant’X by the standardless analysis, and PIXE (Particle Induced X-ray Emission) with linear calibration. A subset of samples was also analyzed by ICP-MS (Inductively Coupled Plasma-Mass Spectrometry). Several metrics including method detection limits (MDLs), precision, bias from a NIST standard reference material (SRM 2783) quoted values, relative absolute difference, orthogonal regression and the ratio of the absolute difference between the methods to claimed uncertainty were used to compare the laboratories. The MDLs were found to be comparable for many elements. Precision estimates were less than 10% for the majority of the elements. Absolute biases from SRM 2783 remained less than 20% for the majority of certified elements. The regression results of PM 10 samples showed that the three XRF laboratories measured very similar mass loadings for S, K, Ti, Mn, Fe, Cu, Br, Sr and Pb with slopes within 20% of unity. The ICP-MS results confirmed the agreement and discrepancies between XRF laboratories for Al, K, Ca, Ti, V, Cu, Sr and Pb. The ICP-MS results are inconsistent with the XRF laboratories for Fe and Zn. The absolute differences between the XRF laboratories generally remained within their claimed uncertainties, showing a pattern generally consistent with the orthogonal regression results.
تدمد: 1352-2310
URL الوصول: https://explore.openaire.eu/search/publication?articleId=doi_________::a5a63a89b6ed3c15600b8db72a110ea9
https://doi.org/10.1016/j.atmosenv.2015.10.084
حقوق: CLOSED
رقم الأكسشن: edsair.doi...........a5a63a89b6ed3c15600b8db72a110ea9
قاعدة البيانات: OpenAIRE