Background: To compare the refractive results of two autorefractors.Methods: We designed this study as an observational, cross-sectional study. We compared the mean spheric and cylinder power, spherical equivalent, Jackson cross-cylinder values, determined the limits of agreement (LoA), and evaluated reliability.Results: Nidek HandyRef-K autorefractor measured more astigmatism and myopia in terms of the mean cylindrical power (PP=.024). Nidek HandyRef-K showed significantly more myopic results in subgroups with spherical power more than +1.50 Diopters (D) hyperopia and -3.00 D myopia (P=.031; P=.045; P=.026; respectively) and higher astigmatism in the subgroup with the cylindric power less than -1.00 D (P. There was a strong correlation for spherical power (Spearman’s rho=0.99, PPPConclusion: Nidek HandyRef-K showed more myopic results regarding spherical equivalent and cylindrical power. Nidek HandyRef-K showed significantly more myopia in spherical power more than +1.50 D hyperopia, -3.00 D myopia, and more prominent astigmatism in cylindrical power less than -1.00 D subgroups.