Off-label use of duct occluder devices to close hemodynamically significant perimembranous ventricular septal defects: A multicenter experience

التفاصيل البيبلوغرافية
العنوان: Off-label use of duct occluder devices to close hemodynamically significant perimembranous ventricular septal defects: A multicenter experience
المؤلفون: Raman Krishna Kumar, Sakshi Sachdev, Rodina Sobhy, Hala Hamza, Atul Kalantre, Floris E.A. Udink ten Cate, Anand Subramanian, Nageswara Rao Koneti, M. Jayranganath, Narayanswami Sreeram
المصدر: Catheterization and Cardiovascular Interventions, 93, 1, pp. 82-88
Catheterization and Cardiovascular Interventions, 93, 82-88
بيانات النشر: Wiley, 2018.
سنة النشر: 2018
مصطلحات موضوعية: Adult, Heart Septal Defects, Ventricular, Male, Cardiac Catheterization, medicine.medical_specialty, Adolescent, Databases, Factual, Septal Occluder Device, Heart block, Alternative therapy, Perimembranous ventricular septal defect, Vascular damage Radboud Institute for Health Sciences [Radboudumc 16], Femoral vein, India, Product Labeling, 030204 cardiovascular system & hematology, Prosthesis Design, Young Adult, 03 medical and health sciences, 0302 clinical medicine, Germany, medicine, Humans, Radiology, Nuclear Medicine and imaging, 030212 general & internal medicine, Child, Retrospective Studies, business.industry, Hemodynamics, Infant, General Medicine, Middle Aged, medicine.disease, Surgery, Treatment Outcome, medicine.anatomical_structure, Child, Preschool, Practice Guidelines as Topic, Feasibility Studies, Egypt, Female, Cardiology and Cardiovascular Medicine, business, Duct (anatomy), Shunt (electrical)
الوصف: Item does not contain fulltext OBJECTIVE: We sought to evaluate the feasibility, technical aspects, and outcome of transcatheter perimembranous ventricular septal defect (pmVSD) closure using duct occluder devices with a single retention disc. BACKGROUND: Use of duct occluder devices to close pmVSD seems a promising alternative therapy. However, limited data exist on this technique. METHODS: From 2010 to 2016, 222 patients (female 47.7%) were identified from databases of five participating institutions in whom pmVSD closure was attempted using an Amplatzer Duct Occluder I or Lifetech duct occluder device. RESULTS: Patients ranged in age from 0.7 to 52 years (median, 7.0 years) and in weight from 4.0 to 70 kg (median, 18.0 kg). The mean size of the VSD was 6.8 +/- 2.2 mm. A large defect (> 6 mm) was present in 137 patients (61.7%). Device closure was successful in 218 patients (98.2%). The 10/8 mm device was used in most patients (n = 85, 38.3%), and the vascular approach was from the femoral vein in 169 patients (76.1%). There were 18 early complications in 17/218 patients (7.8%). Three patients (1.4%) developed complete heart block (transient n = 2; requiring permanent pacing n = 1). Median follow-up was 6 months (6 months-6 years). A mild residual shunt was seen in 10 patients at 6 months follow-up. CONCLUSIONS: The immediate results of transcatheter pmVSD closure using a duct occluder device with a single retention disc are promising. It is an effective technique with a lower rate of complications than for other currently available devices.
تدمد: 1522-1946
URL الوصول: https://explore.openaire.eu/search/publication?articleId=doi_dedup___::09478360429f37ee7cbce1af2186df59
https://doi.org/10.1002/ccd.27792
حقوق: RESTRICTED
رقم الأكسشن: edsair.doi.dedup.....09478360429f37ee7cbce1af2186df59
قاعدة البيانات: OpenAIRE