We have developed a high performance liquid chromatography mass spectrometry method for quantitating paclitaxel and its 6-alpha-OH and 3-para-OH metabolites in 0.1 mL human plasma. After MTBE liquid-liquid extraction, chromatographic separation was achieved with a Phenomenex synergy polar reverse phase (4 μm, 2 mm × 50 mm) column and a gradient of 0.1% formic acid in acetonitrile and water over an 8 min run time. Mass spectrometric detection was performed on an ABI SCIEX 4000Q with electrospray, positive-mode ionization. The assay was linear from 10–10,000 ng/mL for paclitaxel and 1–1,000 ng/mL for both metabolites and proved to be accurate (94.3–110.4%) and precise (