Picosecond X-ray streak camera dynamic range measurement

التفاصيل البيبلوغرافية
العنوان: Picosecond X-ray streak camera dynamic range measurement
المؤلفون: Catherine Goulmy, J. P. Fronty, J. Raimbourg, Patrick Brunel, C. Rubbelynck, D. Gontier, C. Trosseille, C. Zuber, S. Bazzoli
المصدر: The Review of scientific instruments. 87(9)
سنة النشر: 2016
مصطلحات موضوعية: 010302 applied physics, Physics, business.industry, Streak camera, Dynamic range, Streak, Plasma, Electron, 01 natural sciences, 010305 fluids & plasmas, Optics, Picosecond, 0103 physical sciences, business, Instrumentation, Laser Mégajoule, Voltage
الوصف: Streak cameras are widely used to record the spatio-temporal evolution of laser-induced plasma. A prototype of picosecond X-ray streak camera has been developed and tested by Commissariat a l’Energie Atomique et aux Energies Alternatives to answer the Laser MegaJoule specific needs. The dynamic range of this instrument is measured with picosecond X-ray pulses generated by the interaction of a laser beam and a copper target. The required value of 100 is reached only in the configurations combining the slowest sweeping speed and optimization of the streak tube electron throughput by an appropriate choice of high voltages applied to its electrodes.
تدمد: 1089-7623