Segmental Stability and Compressive Strength of Posterior Lumbar Interbody Fusion Implants

التفاصيل البيبلوغرافية
العنوان: Segmental Stability and Compressive Strength of Posterior Lumbar Interbody Fusion Implants
المؤلفون: Seth M. Zeidman, Hani G. Baramki, Thomas Steffen, Anthony Tsantrizos
المصدر: Spine. 25:1899-1907
بيانات النشر: Ovid Technologies (Wolters Kluwer Health), 2000.
سنة النشر: 2000
مصطلحات موضوعية: Joint Instability, Compressive Strength, Rotation, Bone Screws, Models, Biological, Cadaver, Humans, Medicine, Orthopedics and Sports Medicine, Range of Motion, Articular, Intervertebral Disc, Lumbar Vertebrae, business.industry, Neutral zone, Biomechanics, Equipment Design, Anatomy, Initial stability, Middle Aged, Stress shielding, Internal Fixators, Spinal Fusion, medicine.anatomical_structure, Cortical bone, Stress, Mechanical, Neurology (clinical), Implant, business, Cadaveric spasm, Biomedical engineering
الوصف: Study design Human cadaveric study on initial segmental stability and compressive strength of posterior lumbar interbody fusion implants. Objectives To compare the initial segmental stability and compressive strength of a posterior lumbar interbody fusion construct using a new cortical bone spacer machined from allograft to that of titanium threaded and nonthreaded posterior lumbar interbody fusion cages, tested as stand-alone and with supplemental pedicle screw fixation. Summary of background data Cages were introduced to overcome the limitations of conventional allografts. Radiodense cage materials impede radiographic assessment of the fusion, however, and may cause stress shielding of the graft. Methods Multisegmental specimens were tested intact, with posterior lumbar interbody fusion implants inserted into the L4/L5 interbody space and with supplemental pedicle screw fixation. Three posterior lumbar interbody fusion implant constructs (Ray Threaded Fusion Cage, Contact Fusion Cage, and PLIF Allograft Spacer) were tested nondestructively in axial rotation, flexion-extension, and lateral bending. The implant-specimen constructs then were isolated and compressed to failure. Changes in the neutral zone, range of motion, yield strength, and ultimate compressive strength were analyzed. Results None of the stand-alone implant constructs reduced the neutral zone. Supplemental pedicle screw fixation decreased the neutral zone in flexion-extension and lateral bending. Stand-alone implant constructs decreased the range of motion in flexion and lateral bending. Differences in the range of motion between stand-alone cage constructs were found in flexion and extension (marginally significant). Supplemental posterior fixation further decreased the range of motion in all loading directions with no differences between implant constructs. The Contact Fusion Cage and PLIF Allograft Spacer constructs had a higher ultimate compressive strength than the Ray Threaded Fusion Cage. Conclusions The biomechanical data did not suggest any implant construct to behave superiorly either as a stand-alone or with supplemental posterior fixation. The PLIF Allograph Spacer is biomechanically equivalent to titanium cages but is devoid of the deficiencies associated with other cage technologies. Therefore, the PLIF Allograft Spacer is a valid alternative to conventional cages.
تدمد: 0362-2436
URL الوصول: https://explore.openaire.eu/search/publication?articleId=doi_dedup___::385bcbc20aba5ba15db99816675c4348
https://doi.org/10.1097/00007632-200008010-00007
رقم الأكسشن: edsair.doi.dedup.....385bcbc20aba5ba15db99816675c4348
قاعدة البيانات: OpenAIRE