External dosimetry audit for quality assurance of carbon-ion radiation therapy clinical trials

التفاصيل البيبلوغرافية
العنوان: External dosimetry audit for quality assurance of carbon-ion radiation therapy clinical trials
المؤلفون: Tadashi Kamada, Manabu Mizota, Nobuyuki Kanematsu, Tatsuaki Kanai, Shinichi Minohara, Hideyuki Mizuno, Toshihiro Yanou, Shunsuke Yonai, Akifumi Fukumura, Ken Yusa, Masaki Suga, Toshiyuki Shirai
المصدر: Journal of Applied Clinical Medical Physics
بيانات النشر: Wiley, 2018.
سنة النشر: 2018
مصطلحات موضوعية: medicine.medical_specialty, Quality Assurance, Health Care, medicine.medical_treatment, Heavy Ion Radiotherapy, quality assurance, Audit, Radiation Dosage, Imaging phantom, 030218 nuclear medicine & medical imaging, 03 medical and health sciences, 0302 clinical medicine, Neoplasms, Radiation Oncology Physics, Humans, Medicine, Dosimetry, Radiology, Nuclear Medicine and imaging, Medical physics, onsite dosimetry audit, Radiometry, carbon‐ion radiation therapy, Instrumentation, Clinical Trials as Topic, Radiation, Phantoms, Imaging, business.industry, Radiotherapy Planning, Computer-Assisted, Isocenter, clinical trial, multicenter dosimetry, Radiation therapy, Clinical trial, 030220 oncology & carcinogenesis, Ionization chamber, 87.53.Qc, business, Quality assurance, Algorithms
الوصف: Purpose The QA team of the Japan carbon‐ion radiation oncology study group (J‐CROS) was organized in 2015 to enhance confidence in the accuracy of clinical dosimetry and ensure that the facility QA procedures are adequate. The team conducted onsite dosimetry audits in all the carbon‐ion radiation therapy centers in Japan. Materials and Methods A special phantom was fabricated for the onsite dosimetry audit. Target volumes such as the GTV, CTV, and PTV were contoured to the obtained CT images, and two plans with different isocenter depths were created. The dose at the isocenter was measured by an ionization chamber, in the onsite audit and compared with the calculated dose. Results For all the centers, the average of the percentage ratio between the measured and calculated doses (measured/calculated) was 0.5% (−2.7% to +2.6%) and the standard deviation, 1.7%. In all the centers, the beams were within the set tolerance level of 3%. Conclusions The audit demonstrated that the dose at a single point in the water phantom was within tolerance, but it is a big step to say that all doses are correct. In addition, this external dosimetry audit encouraged centers to improve the quality of their dosimetry systems.
تدمد: 1526-9914
URL الوصول: https://explore.openaire.eu/search/publication?articleId=doi_dedup___::4beaaa6a56f9450d0db18be2b221042b
https://doi.org/10.1002/acm2.12465
حقوق: OPEN
رقم الأكسشن: edsair.doi.dedup.....4beaaa6a56f9450d0db18be2b221042b
قاعدة البيانات: OpenAIRE